The door interlock is a switch in normal open condition. When the door is closed and the laser can be
operated, the switch needs to be in closed condition. At this time the interlock at the Laser console is
closed and the Laser can be started. If the door is opened during laser operation, the laser switches off.
________________________________________________________________
When door interlock is not mandated, a connector needs to be in place to defeat the
door interlock in order for the laser to be started. A door interlock plug has to be
used to allow laser operation (P/N 1-0205-007-01-03).
_________________________________________________________________
9.5.
Foot Pedal Connector
The foot pedal is directly attached to the back of the Laser (
Figure 5-1
). The foot pedal can be removed
by pulling the connector from the rear of the laser. The foot pedal is shielded with a protection cover.
SYSTEM OPERATION
10.1.
Switch on Laser System
The Laser console has to be properly seated on a solid base to avoid any movement of the console.
The mains power cord connector is firmly placed into the wall outlet.
The interlock connector is in place and firmly attached.
The foot pedal is in comfortable position.
10.2.
Main Power Switch
The main power switch is located at the back of the Laser console as shown in Figure 10-1. Press the
switch into the “ON” position. When the Laser has been initialized it will go through a series of system-
checks. This allows the internal parameters to be checked and verify the system is running according to
specifications. This routine takes about 2 minutes, and when complete the display will change to the pin
code screen (see
Error! Reference source not found.
).