28
SPECTANO 100 User Manual
DANGER
Standard connection cable is the delivered high-temperature low-noise triaxial cable with a 1:1
connection of the inner conductor, the inner shield (guard) and the outer shield (ground). Connect the
red-labeled cable to the
OUTPUT
connector of
SPECTANO 100
and the blue-labeled cable to the
INPUT
connector of
SPECTANO 100
.
Inner conductor
Inner shield (guard)
Outer shield (ground)
Figure 9-1
Standard high-temperature low-noise triaxial cable
Death or severe injury caused by high voltage or current
►
Before handling
SPECTANO 100
in any way, connect its equipotential
ground terminal with a solid connection of at least 6 mm
2
cross-section and
not longer than 20 meters to the ground terminal in the laboratory.
►
Do not connect or operate
SPECTANO 100
without proper grounding
connection.
►
Before handling
SPECTANO 100
in any way, connect the test object to
ground properly.
►
Do not connect or operate
SPECTANO 100
with a broken, dirty or
oxidized ground terminal or connection cables.
►
Do not use power outlets without protective ground (PE).
►
Only use the delivered AC power supply to power
SPECTANO 100
.