49
SPECTANO Analyzer Suite
10.6.1 Numerical results
The measurement results are displayed numerically on the top of the result view workspace. You can
display the measurement results in frequency domain and in time domain. For information on how to
select the result display, see
10.6.3 "Result view controls".
The following table describes the numerical measurement results in the frequency domain.
Table 10-5
Numerical measurement results in the frequency domain
Data
Description
Tanδ/Power
factor
1
Measured dissipation factor/power factor
ε
r’
Measured real part of the complex relative
permittivity
ε
r’’
Measured imaginary part of the complex
relative permittivity
|Z|
Absolute value of the measured impedance
Phase of Z
Phase of the measured impedance
|Y|
Absolute value of the measured admittance
Phase of Y
Phase of the measured admittance
Cp/Cs
2
Measured parallel/series capacitance
Measured parallel/series resistance
c’
Measured real part of the complex capacitance
c’’
Measured imaginary part of the complex
capacitance
R
Measured resistance
X
Measured reactance
1.
Set by the
Loss angle result
option in the
Options
dialog box (see Figure 10-10 "Options dialog box"
2.
Set by the
Impedance calculation
control (see 10.6.3 "Result view controls"
The following table describes the numerical measurement results in the time domain (PDC).
Table 10-6
Numerical measurement results in the time domain (PDC)
Data
Description
Polarization current
Measured polarization current
Depolarization current
Measured depolarization current
R(t)
Measured electrical resistance over time