The following picture shows a typical transmission calibration setup. The DUT is replaced by a Thru-
connector.
Figure 8-1: Gain, respectively Thru calibration setup in a transmission measurement.
The factory calibration of
Bode 100
moves the reference plane exactly between two cables
of 0.5 m length. So you can measure S21 using the delivered cables having a well calibrated
test-setup.