One-Port measurement example
In this measurement example we will determine the series ans parallel resonance frequency of a 12
MHz quartz filter. Further on, we will display the quartz filter's reflection curve in a Smith chart.
Connect the input
IN 2
of the test object "Quartz Filter" to the OUTPUT of
Bode 100
using a BNC
cable. Further on, connect the BNC short delivered with
Bode 100
to the corresponding output
OUT 2
.
The complete setup is shown in the figure below.
Figure 7-6: Connecting the test object Quartz Filter to
Bode 100
Now start the
Bode Analyzer Suite
and enter the One-Port measurement mode by clicking Impedance
Analysis and then
.
Measurement types and applications
OMICRON Lab
51