
Congratulations
you have successfully used the
B-WIC
impedance measurement adapter with the
Bode 100
. You can load the settings for the measurement by clicking File → Open → and then
navigating to:
"%APPDATA%\OMICRON Lab\Bode Analyzer Suite\Demo Files\"
.
The file you will need is:
ImpAdapt_Inductor.bode3
.
Attention:
Since your setup will be different the file should only be used to explore the different
measurement formats offered by
Bode 100
. Before you execute an own measurement you will have
to
perform a new calibration
.
For a detailed measurement example using the impedance adapter measurement mode to
measure capacitors, please check out the ESR measurement application note at
www.omicron-lab.com/BodeManualAppNotes
7.4.3 Shunt-Thru
The Shunt-Thru measurement mode is especially suitable to measure very small impedance values. It
is basically an S21 transmission measurement in a special configuration. The impedance is calculated
from the measured S21 parameter.
Measurement information
Bode 100
measures
S21
gain and calculates impedance
Z
using the equation:
The Shunt-Thru measurement configuration emulates a 4-wire kelvin connection. The output drives a
current thru the device under test that is shunted between the output signal and GND. Channel 2
measures the voltage drop that occurs at the DUT. This measurement mode offers highest sensitivity
for low-impedance value DUT. The recommended impedance measurement range starts at roughly
1 mΩ. With amplifiers or pre-amplifiers also µΩ measurements can be performed.
Shunt-Thru measurement configuration offers low sensitivity for high impedance DUTs. It is not
recommended to measure impedance values much higher than 100 Ω.
The Shunt-Thru configuration inherently suffers a ground-loop error at low frequencies. The
current flowing thru the cable shield of the connection to Channel 2 ground introduces a
measurement error that can become significant at frequencies below 10 kHz when
measuring very low impedance values. To suppress respectively reduce the ground-loop
error at low frequencies, use a common-mode choke or common-mode transformer.
64
OMICRON Lab
Summary of Contents for Bode 100
Page 1: ...Smart Measurement Solutions Bode 100 User Manual...
Page 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Page 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Page 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...