17 |
P a g e
5.2.2.5
Smart Sensor
Refer to the
Smart Sensor Device Interface
documentation.
5.2.3
RTD Configuration
RTD Configuration
7
6
5
4
3
2
1
0
Available
Assigned
Apply
Scaling
Lock
Sensor Type (Range)
0
*
?
?
See Below
5.2.3.1
Sensor Type / Range
Sensor Type
Sensor Input Type (Range)
Measurement Type
0x00
100 Ohm, 385 Curve
0x21 Temperature (°C)
0x01
500 Ohm, 385 Curve
0x21 Temperature (°C)
0x02
1000 Ohm, 385 Curve
0x21 Temperature (°C)
0x03
100 Ohm, 392 Curve
0x21 Temperature (°C)
0x04
100 Ohm, 3916 Curve
0x21 Temperature (°C)
5.2.3.2
Apply Scaling
If set, the user defined Offset and Gain values will be used to adjust the sensor reading:
Result = (Raw Reading * Gain) + Offset
5.2.3.3
Lock
If set, the user specified units of measure string (4-character maximum) will be used in place of the default units of
measure.
5.2.3.4
Assigned
Refer to the
Smart Sensor Device Interface
documentation.
5.2.3.5
Available
Refer to the
Smart Sensor Device Interface
documentation.
5.2.4
RTD Input Sensor Device
The Sensor Device field determines the connector signal configuration.
I/O (RTD)
7 6
5
4
3
2
1
0
Description
X X
X
X
X
X
0
0
Not Available/Used
X X
X
X
X
X
0
1
2 Wire, configuration
X X
X
X
X
X
1
0
3 Wire, configuration
X X
X
X
X
X
1
1
4 Wire, configuration
5.2.5
RTD Input Connector
A (bit 6)
Description
M12-4 Connector Pins
1
2
3
4
0x00
Wiring Option #1
S+
P+
P-
S-
0x01
Wiring Option #2
S+
S-
P-
P+