2-3
EN
FIGURE 2-2. HART Communicator ScreenWithout Model PX2088 Smart Device Driver.
REVIEW CONFIGURATION DATA
Review all of the factory-set configuration data to ensure that is reflects the needs of
your application before operating the transmitter in an actual installation.
Review
Review the transmitter configuration parameters set at the factory to ensure
accuracy and compatibility with your particular application. After activating the
review function, scroll through the data list to check each variable. Refer to “Basic
Setup” in this section of the manual if a change to the transmitter configuration data
is necessary.
CHECK OUTPUT
Before performing other transmitter on-line operations, review the digital output
parameters to ensure that the transmitter is operating properly and is configured to
the appropriate process variables.
Process Variables
The process variables for the Model PX2088 Smart provide the transmitter output,
and are continuously updated. The Process Variable menu displays the following
process variables:
n
Pressure
n
Percent Range
n
Analog Output
HART Fast Keys
1, 5
HART Fast Keys
1, 1
Generic
Online
1 Device Setup
2 PV 100.00 inH2O
3 AO 20.00 mA
4 LRV 0.00 inH2O
5 URV 100.00 inH2O