Figure 10. General Dimensions, OS4000-WC Water Cool Jacket
Figure 11. General Dimensions, Vacuum Bushing OS4000-VB
The following steps outline how to install & operate the fiber optic transmitter:
• Install the fiber optic lens probe
• Connect the fiber optic cable to the transmitter front plate. Figure 8 shows the
fiber FC connector (keyed) to the mating transmitter connector. There are two
ports for connections. The temperature port is for measuring temperature. The
laser sighting port is for aligning & positioning the fiber optic lens probe.
• Connect the fiber optic cable to the laser sighting port.
• Connect the power/output cable connector to the mating connector on the
back plate of the transmitter.
• Power the transmitter using a DC power supply. Table 1 (page 2-6) & Figure 6
(page 2-3) show the wiring connections.
2.5-16 UNF-2A
OUTLET
INLET
APPLY
CLEAN AIR
BUILT-IN
AIR PURGE
COLLAR
BARB ADAPTER 3/16" ID
1/4-28 UNF THREAD
2.5-16 UNF-2B
146.05 (5.75)
DEPTH OF ENGAGEMENT OF OS4001
8.89 (0.35)
21.59
(0.85)
ø 82.55
(3.250)
DIMENSIONS mm (in)
ø = DIAMETER
3/8-24
THREAD
6-32 THRD
2 PLCS
17.45
(0.687)
“D”
19.05
(0.75)
42.93
(1.69)
ø 5.33
(0.210)
DIMENSIONS mm (in)
ø = DIAMETER
VACUUM
TO LENS
PROBE
ATMOSPHERE
TO FIBER OPTIC
TRANSMITTER
Ø D = 0.500" / 0.750" / 1.000"
2-5
Installation - Laser Sighting
2