background image

 

UNRESTRICTED 

INTERFERENCE FROM MOVEMENT (OS-MINI) 

The low-noise sensing head cable on -HA models is resistant to interference caused by movement. The 

sensing head may be mounted on moving machinery such as robot arms without affecting the accuracy of 
the measured temperature.

 

ELECTRICAL INTERFERENCE 

The OS-MINI and OS-MINI22 are tested to industrial standards for electromagnetic compatibility (EMC) 
as shown in Specifications at the beginning of this manual.  

To minimize electromagnetic interference or ‘noise’, the OS-MINI series sensor should be mounted away 
from motors, generators and such like. 

POWER SUPPLY (OS-MINI and OS-MINI22) 

Be sure to use a 24 V DC (100 mA) power supply. 

MECHANICAL INSTALLATION 

Both OS-MINI and OS-MINI22 sensors come with a 1 m cable and a mounting nut as standard, and the 
OS-MINIUSB comes with a 1.5m cable and mounting nuts. Longer cables are available to order. The 
sensor can be mounted on brackets or cut-outs of your own design, or you can use the fixed or adjustable 
mounting bracket accessories.  

Note for OS-MINI and OS-MINI22:

 The sensor housing must be connected to earth at one point, either 

the housing of the sensing head, the electronics module, or the output cable shield termination. To avoid 
ground loops, please ensure the sensor is grounded at only one of these points.  

ELECTRICAL INSTALLATION 

OS-MINI AND OS-MINI22 WIRING (ALL MODELS) 

Check the distance between the sensing head and the electronics module, and between the electronics 
module and the instrumentation. If necessary, the sensor can be ordered with a longer cable between the 
sensing head and the electronics module. 

The output cable from the electronics module should have an outer diameter between 3.0 and 6.5 mm, 
with conductors of size 28 to 18 AWG.  

The terminal blocks in the electronics module may be removed from the circuit board for easy wiring. 

Do not disconnect the touch screen circuit board from the main circuit board while the sensor is on. 

WIRING (-C4, -C4-R-D, -BB AND -BRT MODELS)  

When connecting several sensors in a single Modbus network, all of the sensors should be connected via 
a junction box to a single network bus cable, running from the furthest sensor to the Modbus Master. 

Up to 247 sensors may be connected to a single Modbus network. Each sensor must have a unique 
Modbus address. OS-MINI series sensors are normally shipped with Modbus address 1. The Modbus 
address may be changed via the touch screen interface on -BRT models, or via Modbus. 

To help prevent data reflections, please ensure the cable between each sensor and the main network bus 
is as short as possible. The network bus should be terminated with a resistor of 120

Ω between the RS+ 

and RS- wires. The PWR- wire of the bus should be connected to the signal ground of the Modbus 
Master.  

 

Summary of Contents for OS-MINI SERIES

Page 1: ...e mail info omega com For latest product manuals omega com en us pdf manuals TM Shop online at omega com User sGuide Miniature Fixed Infrared Sensor with Optional Touch Screen Display...

Page 2: ...tice Servicing North America U S A Omega Engineering Inc Headquarters 800 Connecticut Ave Suite 5N01 Norwalk CT 06854 Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canad...

Page 3: ...s such as robot arms Different choices of optics are available to measure small or large targets at short or long distances and there is a choice of 4 20 mA RS485 Modbus and alarm relay outputs OS MIN...

Page 4: ...RoHS Compliant Yes Yes Yes ELECTROMAGNETIC COMPATIBILITY STANDARDS Class Standard Description EMC Directive EN61326 1 2006 Electrical equipment for measurement control and laboratory use Industrial Im...

Page 5: ...C The high ambient temperature sensing head on HA models is capable of withstanding temperatures of up to 180 C without cooling It is available with 20 1 optics There is no need to supply cooling air...

Page 6: ...2 C or 1 of reading whichever is greater Repeatability 0 5 C or 0 5 of reading whichever is greater Emissivity Setting Range 0 10 to 1 00 Emissivity Setting Method CB models via two rotary switches i...

Page 7: ...n condensing Maximum 95 non condensing CE Marked Yes Yes Yes RoHS Compliant Yes Yes Yes ELECTROMAGNETIC COMPATIBILITY STANDARDS The OS MINI22 series conforms to EMC Directive EN61326 1 2006 Electrical...

Page 8: ...tput and Interface OS MINI MA 0 C to 60 C 302 30 1 divergent CT configurable within the limits 20 to 1000 C C4 R D RS485 Modbus output and two alarm relay outputs with touch screen EXAMPLE OS MINI22 S...

Page 9: ...lectronics module provides a large bright display of the measured temperature as well as options for full configuration of the sensor The graph view shows the history of the measured temperature In al...

Page 10: ...y 0 2 to 1 0 Response Time t90 125 ms 90 response Spectral Range 8 to 14 m Supply Voltage 5 V DC provided by USB Supply Current 50 mA max VIRTUAL COM PORT Baud Rate 9600 baud Format 8 data bits no par...

Page 11: ...MBERS Field of View Model Number 2 1 OS MINIUSB SN21 20 1 OS MINIUSB SN201 OPTICAL CHART The optical chart below indicates the nominal target spot diameter at any given distance from the sensing head...

Page 12: ...ssed measured temperature MicroSD Card status This icon is displayed when an SD card is inserted and flashes when data logging is in progress This icon is displayed when scheduled data logging is enab...

Page 13: ...wards and forwards in time touch the graph and drag it The graph stores the most recent 24 hours of temperature data Reset Graph Clears and restarts the graph Real Time Scrolling View Returns the grap...

Page 14: ...temperature it holds this temperature steady for the Hold Period Valley Similar to Peak Hold mode except that the sensor outputs the minimum detected temperature steady for the Hold Period Off Disable...

Page 15: ...mA Temperature at 20 mA Please note The lower temperature range limit Minimum 20 C Maximum 900 C The upper temperature range limit Minimum 80 C Maximum 1000 C The difference between the temperatures...

Page 16: ...e drops below the Alarm Set Point Off The alarm is disabled Automatic The alarm is acknowledged and reset automatically when the temperature reaches the reset temperature see Hysteresis It can also be...

Page 17: ...optional battery is fitted DATA LOGGING SPECIFICATIONS Data Logging Interval 1 to 86 400 seconds 1 day MicroSD Card Max capacity 2 GB not included Internal Clock Battery 1 x BR 1225 3V not included Va...

Page 18: ...e not compatible with the OS MINI22 INSTALLATION OF MICROSD CARD AND BATTERY DATA LOG FILES Data is saved to the MicroSD Card in csv format This file format can be opened or imported by spreadsheet so...

Page 19: ...n Setup exe 3 If you downloaded the software unzip the files and run Setup exe 4 Follow the on screen instructions Using the Software for the First Time 1 After the software has been installed when pr...

Page 20: ...ut emissivity contact Omega Reflected Energy Compensation In most applications the target surface has the same surroundings as the sensor for example it is in the same room In this case Reflected Ener...

Page 21: ...Temperature R 0x11 1 Status bits active high Bit 0 Measurement error Bit 1 Sensor temperature low Bit 2 Sensor temperature high Bit 3 Object temperature low Bit 4 Object temperature high R 0x16 1 Ref...

Page 22: ...INI Removable spigot fitted to type OS210 APSN for use with 151 201 301 and CF models Not fitted to type OS210 APSW for 21 models For OS MINIUSB Removable spigot fitted to type OS210 APSN for use with...

Page 23: ...ctive plastic window with stainless steel holder When screwed over the lens of the sensor this provides extra mechanical protection for the germanium lens and helps retain fragments of the lens in cas...

Page 24: ...to the lens The use of shields may help in this respect For further information and assistance contact Omega AMBIENT TEMPERATURE The OS MINI is available with a choice of two miniature sensing heads f...

Page 25: ...only one of these points ELECTRICAL INSTALLATION OS MINI AND OS MINI22 WIRING ALL MODELS Check the distance between the sensing head and the electronics module and between the electronics module and...

Page 26: ...UNRESTRICTED WIRING DIAGRAMS OS MINI AND OS MINI22...

Page 27: ...UNRESTRICTED...

Page 28: ...mber Bits 20 23 Sensor type 12 OSMini Bits 24 26 Sensor field of view For MA 0 2 1 1 15 1 2 30 1 For HA 0 20 1 Bits 28 32 Reserved R 0x06 1 Unfiltered object temperature R 0x08 1 Sensor temperature R...

Page 29: ...reset 1 manual reset 0 R W MW Bit 4 Alarm acknowledge R W MW Bit 5 Alarm reset R W MW Bits 6 7 Reserved Bit 8 High alarm 1 low alarm 0 R W MW Bit 9 Alarm enabled 1 disabled 0 Bits 10 15 Reserved R W...

Page 30: ...electromagnetic fields e g around arc welders or induction heaters Electromagnetic interference can cause measurement errors Wires must be connected only to the appropriate terminals VIEWING THROUGH A...

Page 31: ...lect Run as Administrator Do not connect the sensor until prompted by the installer Software will not start or will not install Incompatible Windows version Try installing on a PC with Windows Vista 7...

Page 32: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Page 33: ...Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Indus...

Reviews: