UNRESTRICTED
Emissivity and Compensation
Emissivity
Setting
Enable
Reflected
Energy
Compensation
Reflected
Temperature
Enter the emissivity of the target. Target emissivity can be determined experimentally
or estimated using an emissivity table. For more information, contact Calex. Minimum:
0.2. Maximum: 1.0.
If enabled, compensates for errors caused by reflected energy from hotter or colder
objects.
Enter the temperature of the surroundings of the target for Reflected Energy
Compensation. Minimum: -20°C. Maximum: 1000°C.
4 to 20 mA Output (MA-R-D models)
Set the temperature range limits for the 4 to 20 mA output.
Temperature
at 4 mA
Temperature
at 20 mA
Please note
The lower temperature range limit. Minimum: -20°C. Maximum: 900°C.
The upper temperature range limit. Minimum: 80°C. Maximum: 1000°C.
The difference between the temperatures at 4 mA and at 20 mA must be at least
100°C. The temperature at 20 mA must be greater than the temperature at 4 mA.
Modbus Address (C4-R-D models)
Modbus
Address
The current Modbus address of the sensor is displayed. Enter a new address, then
press Apply to save it to the sensor. Cycle the power to use the new address.
Minimum: 1. Maximum: 247.
Alarms
Configure the settings for Alarm 1 and Alarm 2 separately and configure alarm logging
settings.
Manually Reset Alarms
If an alarm has been triggered, allows both alarms to be triggered again. Alarms will not
be triggered again until they are reset, either automatically or manually.