background image

AC Voltage Data Logger

9

B

ATTERY 

R

EPLACEMENT

The OM-DLAV Data Logger uses a 9-Volt Alkaline battery which will have an av-
erage life of 1 year. The actual life of the battery will depend on a variety of fac-

tors such as the temperature the battery is used in, the sampling rate of the log-
ger, and how often the OM-DLAV is used. Extreme temperatures will result in
reduced battery life. If the battery is exhausted during a recording, the OM-DLAV
will automatically terminate the recording and save any data recorded up to that
point. The LED will stop flashing and the OM-DLAV will not allow a new record-
ing to be started until the battery has been replaced. Recorded data can be re-
trieved regardless of the condition of the battery.

When the logger will not be used for an extended period, or when the battery

is exhausted, it is strongly recommended to remove the battery to avoid dam-

age to the logger from battery leakage. 

Use only Alkaline batteries in the

OM-DLAV. Substitution of a different battery can result in reduced

performance or permanent damage to this instrument.

To replace the OM-DLAV’s battery follow the steps below:

1.

 

Make sure the logger is not set in the recording position and is dis-
connected from any voltage source. If the logger is recording slide the door
open. Any information recorded up to this point will not be lost and is auto-
matically saved.

2.

 

Remove the rear battery cover.

3.

 

The battery and battery clip may be extended from the battery compartment
to facilitate replacement.

4.

 

Remove the battery from the logger and dispose of properly. To assure
proper reset of the logger, wait at least one minute before connecting new

battery.

5.

 

Line up the connector of the new battery and push onto the pins of the bat-
tery clip with a firm motion. 

Do NOT reverse the battery connection, even

momentarily, as this can result in loss of data or permanent damage to
the logger

.

6.

 

Place the battery in the battery compartment of the OM-DLAV.

7.

 

Replace the battery cover.

Once the battery has been replaced, it will be necessary to reset the logger’s
internal clock. Refer to the section, “Setting the Data Logger” for instructions on
how to do this.

Summary of Contents for OM-DLAV

Page 1: ...AC Voltage Data Logger 1...

Page 2: ...r 4 If you are using Windows 3 1 click on Program Manager 5 Select FILE and select RUN 6 Type A SETUP and press Enter To get the latest product information and current release of the Data Logger to PC...

Page 3: ...s are indicated in parentheses next to the recording interval The recording intervals that correspond to charts used by the OM DP3 Data Plotter show the appropriate chart selection instead of a maximu...

Page 4: ...s reached its maximum capacity of 10 000 samples NOTE If the battery is exhausted during a recording the OM DLAV will automatically terminate the recording and save any data recorded up to that point...

Page 5: ...wer is off and the equipment is safe to touch Always remove the OM DLAV leads before connecting the OM DLAV to your computer RETRIEVING AC VOLTAGE RECORDINGS 1 Disconnect the OM DLAV from the AC volta...

Page 6: ...ly be displayed in the most re cently used mode To view a recording as a graph select the graph option in the Dis play Format box The pro gram will display the entire recording as a plotted graph with...

Page 7: ...our mind VIEW A RECORDING AS TEXT 1 When a recording is retrieved or when a saved recording is opened it is displayed in the most recently used mode The recording can be viewed as a text listing or a...

Page 8: ...sword in the Current Password section If you want to have a new password next time you can type a new password in the New Password section Click the OK button when finished If this is the first time y...

Page 9: ...n of a different battery can result in reduced performance or permanent damage to this instrument To replace the OM DLAV s battery follow the steps below 1 Make sure the logger is not set in the recor...

Page 10: ...Battery Life 1 Year Average OM DP3 DATA PLOTTER The OM DP3 is an optional accessory instrument that allows the user to retrieve data and program data loggers in the field without the need of a comput...

Reviews: