5
Product Maintenance
Battery Replacement
Materials:
Small Phillips Head Screwdriver & Replacement Battery (OM-CP-BAT105)
1. Puncture the center of the back label with the screw driver and unscrew the enclosure.
2. Remove the battery by pulling it perpendicular to the circuit board.
3. Insert the new battery into the terminals and verify it is secure.
4. Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
The OM-CP-VOLT101A and OM-CP-VOLT101A-160MV standard calibration depends on the range.
Voltage
160mV
2.5V
15V
30V
Range
0V and 90-160mV
0V and 2.25-2.5V
0V and 13.5-15V
0V and 27-30V
Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in
the software when the device is due.
OM-CP-VOLT101A & OM-CP-VOLT101A-160MV