Product Maintenance

Battery Replacement

Materials:

 Small Phillips Head Screwdriver & Replacement Battery (OM-CP-BAT105)

1.  Puncture the center of the back label with the screw driver and unscrew the enclosure.
2.  Remove the battery by pulling it perpendicular to the circuit board.
3.  Insert the new battery into the terminals and verify it is secure.
4.  Screw the enclosure back together securely.

 

Note: Be sure not to over tighten the screws or strip the threads.

Recalibration 

The OM-CP-VOLT101A and OM-CP-VOLT101A-160MV standard calibration depends on the range.

Voltage

160mV

2.5V

15V

30V

Range

0V and 90-160mV

0V and 2.25-2.5V

0V and 13.5-15V

0V and 27-30V

Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in 
the software when the device is due.

OM-CP-VOLT101A & OM-CP-VOLT101A-160MV

Summary of Contents for OM-CP-VOLT101A

Page 1: ...M0000 0016 INSTRUCTION SHEET Shop online at omega comSM e mail info omega com For latest product manuals www omegamanual info Xxxxx Xxxxxx Xxxxx Xxxxxx Xxxxxxxx M0000 0016 INSTRUCTION SHEET Shop onlin...

Page 2: ...15V and 30V models Single Ended Wiring OM CP VOLT101A 160MVDC Differential Wiring Two position removable screw terminal connections accepts 2 wire configurations Three position removable screw termina...

Page 3: ...be that the OM CP IFC200 is not properly connected 1 In the software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all...

Page 4: ...rs appropriate for the data logging application and click Start Quick Start applies the most recent custom start options Batch Start is used for managing multiple loggers at once Real Time Start store...

Page 5: ...inals and verify it is secure 4 Screw the enclosure back together securely Note Be sure not to over tighten the screws or strip the threads Recalibration The OM CP VOLT101A and OM CP VOLT101A 160MV st...

Page 6: ...FC200 Baud Rate 115 200 Typical Battery Life 10 years at a 15 minute reading rate Operating Environment 40 C to 80 C 40 F to 176 F 0 RH to 95 RH non condensing Enclosure Material ABS plastic Dimension...

Page 7: ...e VDC 160mVDC Measurement Resolution 5 V Calibrated Accuracy 0 01 FSR Memory 1 000 000 readings software configurable memory wrap 330 000 in enhanced start stop mode Reading Rate 4 readings very secon...

Page 8: ...a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans Should any Product s be used in or with any nuclear installat...

Reviews: