background image

5

OM-CP-ULTRASHOCK-A & OM-CP-SHOCK300  

Humidity

The OM-CP-ULTRASHOCK-A utilizes polymer type humidity sensor and a high-tech Gore® vent material at 
the entrance holes to the internal temperature and humidity sensor chamber. The vent equalizes pressure, 
reduces condensation and also blocks entry of liquids and other contaminants.

If the unit has been exposed to very high levels of humidity for extended time, it will return to normal 
measurement accuracy by storing it in normal room humidity (approximately 50% RH) conditions for several 
days.

Temperature

The OM-CP-ULTRASHOCK-A utilizes a silicon temperature sensor and is calibrated for accuracy. When 
measuring humidity and temperature, in addition to shock, please note that there will be a delay in response 
time to humidity and temperature measurements as the OM-CP-ULTRASHOCK-A processes shock events 
before other parameters.

Pressure

The OM-CP-ULTRASHOCK-A pressure sensor is designed to measure barometric pressure from very low 
altitudes to those altitudes found in aircraft, up to high altitude balloons. While specified to 300 mbar, it will 
continue to report data to 100 mbar. This equates to approximately 53,000 feet.

The pressure sensor is contained within the same sealed chamber as the temperature and humidity sensors 
and is also protected by the Gore® vent.

Device Operation

Enabling and Disabling of Channels

1. Connect the logger to the interface cable.
2. Highlight the data logger in the Connected Devices 

list.

3. Click 

Properties

 or right click the logger in the list 

and select 

Properties

.

4. Select 

Channels

 in the Properties list.

5. Check or uncheck 

Enable data recording

 for each 

channel you want to enable or disable.

Summary of Contents for OM-CP-ULTRASHOCK-A

Page 1: ...nd Tri Axial Shock Data Logger OM CP SHOCK300 Tri Axial Shock Data Logger MQS5089 0120 Xxxxx Xxxxxx Xxxxx Xxxxxx Xxxxxxxx M0000 0016 INSTRUCTION SHEET Shop online at omega comSM e mail info omega com...

Page 2: ...ies the most recent custom start options Batch Start is used for managing multiple loggers at once Real Time Start stores the dataset as it records while connected to the logger The status of the devi...

Page 3: ...accelerometer 100g and 300g high g accelerometer Physical Mounting The most accurate transfer of shock vibration and acceleration to the final logged data will occur when the mounting is mechanically...

Page 4: ...out USB connected and adequate battery charge level Flashes every 15 seconds maximum Indicates low battery charge level Logging Data with USB communication Flashes every 15 seconds Indicates normal op...

Page 5: ...lay in response time to humidity and temperature measurements as the OM CP ULTRASHOCK A processes shock events before other parameters Pressure The OM CP ULTRASHOCK A pressure sensor is designed to me...

Page 6: ...nly sample data at the logging rate For instance if a 30 second logging rate was selected then only a single data point is captured at each 30 second instant The potential to lose valuable data can be...

Page 7: ...red at the selected logging rate The trigger sample count should be chosen carefully because it may gather data points for which we have no interest For instance if 100 were entered and the logging ra...

Page 8: ...only used for the duration of the triggered shock events plus a post event buffer which is user programmable Product Maintenance Battery Charging and Care The OM CP ULTRASHOCK A and OM CP SHOCK300 bat...

Page 9: ...midity sampled approx every 2 seconds at intervals shorter than 2 seconds Otherwise sampled at the reading rate GENERAL Memory 3 593 358 readings 598 893 per channel all channels in use Start Modes So...

Page 10: ...l Time Recording May be used with PC to monitor and record data in real time at a 1 second sample rate Password Protection An optional password may be programmed into the device to restrict access to...

Page 11: ...a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications or used on humans Should any Product s be used in or with any nuclear installati...

Reviews: