Device Maintenance
Battery Replacement
Materials: 3/32” HEX Driver (Allen Key) and Replacement Battery (OM-CP-BAT103)
- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
The OM-CP-QUADVOLT or OM-CP-OCTVOLT standard calibration is dependant upon the range.
Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in
the software when the device is due. Specifications subject to change. See Omega’s terms and conditions at
www.omega.com
OM-CP-QUADVOLT and OM-CP-OCTVOLT
Range
100mV
2.5V
15V
30V
Calibration Point
0mV and
90-100mV
0mV and
2.25-2.5V
0mV and
14.9-15.5V
0mV and
27-30V