Device Maintenance

Battery Replacement

Materials: 3/32” HEX Driver (Allen Key) and Replacement Battery (OM-CP-BAT103)
- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together securely.

Note: Be sure not to over tighten the screws or strip the threads.

Recalibration 

The OM-CP-QUADVOLT or OM-CP-OCTVOLT standard calibration is dependant upon the range. 

Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in 
the software when the device is due. Specifications subject to change. See Omega’s terms and conditions at 
www.omega.com

OM-CP-QUADVOLT and OM-CP-OCTVOLT

Range

100mV

2.5V

15V

30V

Calibration Point

0mV and

90-100mV

0mV and

2.25-2.5V

0mV and

14.9-15.5V

0mV and

27-30V