d. INPUT RANGE BIT NO.
VALUE
FREQUENCY/PULSE (FP)
0
0
----
1
Low input level
1
0
----
1
Debounce contact
2
0
----
1
3K pull up to 5V
3
0
----
1
1K pull down
5-4
00
12.5V excit.
01
5V excit.
10
8V excit.
11
not used
6
0
Disable Reading Scale and Offset
1
Enable Reading Scale and Offset
7
not used
2. INPUT/OUTPUT CONFIGURATION:
This is one byte data (2 characters).
Bit pattern is:
BIT NO. VALUE
MEANING
TC/RTD
ACV/ACC/ST
PR
FP
1-0
00
Degree C
----
----
01
Degree F
----
Freq. mode
10
Degree K
----
Totalizer
----
11
Degree K
----
Totalizer
Freq. mode
2
0
Temp. compensation ----
See Note 1
----
1
No temp compensation ----
See Note 1
Quadrature
3
0
----
----
See Note 1
----
1
----
----
See Note 1
A-B mode
4
0
----
----
----
----
1
----
----
----
Totalize mode
5
0
----
----
----
----
1
----
----
Square root
----
7-6
not used set to 0
Note 1:
Bit 2&3 set the speed of process totalize. The time it would take process totalize to
reach the non-totalize reading is shown in the following table:
Page 10