![background image](http://html2.mh-extra.com/html/omega/hhm9007r/hhm9007r_user-manual_4023396017.webp)
15
Capacitance (6000 counts)
Range
Resolution
Accuracy
6nF
0.001nF
± (3.0% red+30d)
60nF
0.01nF
± (3.0% red+10d)
600nF
0.1nF
6µF
0.001µF
60µF
0.01µF
600µF
0.1µF
6mF
1µF
± (5.0% red+10d)
Overload protection: 600V DC or 600V AC RMS
Temperature
Range
Resol
-ution
Accuracy
Sensor
type
0°C ~ 400°C
0.1°C
± (1.0%
red+1°C )
K-type
Thermocou-
ple
-50°C ~ 0°C,
400°C ~ 1300°C
± (2.0% red+3°C)
32°F ~ 750°F
0.1°F
± (1.0% red+2°F)
-58°F ~ 32°F,
750°F ~ 2372°F
± (2.0% red+6°F)
Overload protection: 30V DC or 30V AC RMS
% Duty Cycle
Range
Resolu-
tion
Pulse
Width
Accuracy
(5V logic)
5% to 95%
0.1%
> 10µs
± (2.0%
red+10d )
Frequency range: 5% to 95% (40Hz to 1kHz),
10% to 90% (1kHz to 10kHz), 20% to 80% (10kHz to
20kHz)
Overload protection: 600V DC or 600V AC RMS