background image

HFS Heat Flux Sensor Instruction Manual 

 

 

 

Take the sensor’s temperature measurement, T

°C

,  at  each  measurement  in  time 

along  with  the  calibrated  sensitivity,  S

Calib

,  to  determine  the  heat  flux  sensitivity  at  that 

specific temperature, S

@ T

°C

.  

 

𝑆

𝑇℃

 = 

𝑀

𝐹.∗ 𝑆

𝐶𝑎𝑙𝑖𝑏

 

= [0.00334 ∗ 𝑇

℃ 

+ 0.917 ]  ∗ 𝑆

𝐶𝑎𝑙𝑖𝑏

 

 

 

 

 

 

Calculating Heat Flux

 

 

Using the DC voltage measurements taken from the heat flux wire leads (white & 

bright red), the heat flux values can be calculated using the adjusted sensitivity for that 
specific sensor and the following equation.  

 

 

 

 

where 

q”

 is the heat flux absorbed through the sensor, 

∆V

q”

 is the HFS sensor output heat 

flux voltage, and S

@ T

°C

. is the sensitivity of the sensor that has been adjusted according 

to the sensor’s temperature at that specific time. 

 

 

 

 

For example:

 A voltage value of 1.80 mV is measured across the heat flux  

leads, and the sensor’s calibrated sensitivity is specified to equal 0.90 µV/(W/m

2

), 

and  the  sensor’s  temperature  is  measured  to  be  equal  to  30°C  from  the 
thermocouple leads at that point in time. The calculation for heat flux is as follows.  

1. 

First, adjust the sensitivity according to the measured sensor temperature 
from the sensor’s integrated Type-T thermocouple. (Ignore this step if using 
the UHFS-09 sensor).  

 

 

  

M.F. = 0.00334*T

 

°

 

 

+ 0.917

 

 

 

0.5

 

0.6

 

0.7

 

0.8

 

0.9

 

1

 

1.1

 

1.2

 

1.3

 

1.4

 

-50

 

-30

 

-10

 

10

 

30

 

50

 

70

 

90

 

110

 

130

 

Sensor Temperature (°C)

   

 

 

Sensitivity Multiplication Factor Versus 

 

Sensor Temperature (25

 

°

 

C Origin)

 

 

 

Summary of Contents for HFS-5

Page 1: ...sensor temperature measurements The sensitivity of each sensor is provided with each unit recorded on their respective calibration certificate Sensor calibration procedures adhere to ASTM standard C11...

Page 2: ...Heat Flux Sensor Sensitivity S V BTU ft2 hr V W m2 Temperature T F C Temperature Difference T F C Heat Flux Sensor Sensitivity at a Temperature S T C V BTU ft2 hr V W m2 Sensitivity Multiplication Fac...

Page 3: ...tion to the temperature dependence function 6 Calculate heat flux using the adjusted sensitivity 7 Remove the sensor from the measurement surface if necessary while being very careful not to damage it...

Page 4: ...ux Sensor Connect an ohmmeter to the bright red wire and the white wire leads to check the heat flux sensor s electrical resistance It should 1000 for HFS 5 sensors or 5 k for HFS 6 UHFS 09 sensors If...

Page 5: ...ial number should be located on a tag on the wire leads Sensor Mounting Installation The manner in which the heat flux sensor is mounted depends on the application that it is being used for The best r...

Page 6: ...r is removed from them IMPORTANT When removing the sensor very carefully remove the side with the leads with one hand and peel the opposite side with the other hand to avoid bending as much as possibl...

Page 7: ...S T C is the sensitivity of the sensor that has been adjusted according to the sensor s temperature at that specific time For example A voltage value of 1 80 mV is measured across the heat flux leads...

Page 8: ...on The sensor sensitivity is the output voltage induced by the sensor divided by the heat flux conducted through the sensor Using a custom made calibration apparatus Heat flux can be calculated using...

Page 9: ...teel containing up to 0 35 lead by weight Exemption 6 b Lead as an alloying element in aluminum containing up to 0 4 lead by weight Exemption 6 c Copper alloy containing up to 4 lead by weight and Exe...

Page 10: ...echa europa eu chem_data authorisation_process candidate_list_table_en asp All HFS 5 HFS 6 UHFS 09 heat flux sensor products will have the following REACH Compliance REACH Status Compliant The REACH C...

Page 11: ...belongs to the following product Objects of the Declaration Product Model Number HFS 5 HFS 6 UHFS 09 The object of the declaration described above is in conformity with the relevant Union harmonizati...

Reviews: