background image

FTB700-T INSTRUCTIONS

 

Page 9 

omega.com

SETTINGS AND CALIBRATION

Setting Frequency

The  FTB700-T  converts  a  train  of  off/on  pulses  from  the 

flow sensor into a continuous milliAmp signal that ranges 

from 4 mA at zero flow to 20 mA at the desired maximum 

flow. The desired maximum is determined by the user and 

entered as a frequency as follows:

1. 

Decide what flow rate should represent the top of 

the scale. This is ordinarily the maximum expected 

flow, or a value just above it, in gallons per minute.

2.  Locate the K-factor of the flow sensor (found on 

the meter or fitting. The K-factor is the number of 

pulses the flow sensor produces per gallon of flow.

3. 

Calculate frequency, using this formula:

K-Factor  x  

Top Flow (GPM)

 = Frequency 

                             60

4. 

Enter the frequency using the four rotary Frequency 

switches. Note the decimal point between the third 

and fourth switches.

Setting Averaging Time

For  most  applications,  this  step  can  be  ignored,  as  the 

standard setting will work fine. However, when a particularly 

steady output signal is desired, or in large pipe, a larger 

averaging  period  may  be  desirable.  Note  however  that 

the  averaging  period  requires  a  tradeoff,  since  a  longer 

averaging period implies a slower response time. If steady 

signal is more important than fast response, increase the 

averaging time as desired. See the diagram below for the 

switch positions and their corresponding times.

Checking Calibration

Normally it should not be necessary to check calibration, 

since the digital design of this unit virtually eliminates drift. 

However, there are two types of calibration check that can 

be performed. Look at the diagram below to locate the 4 

and 20 mA force switches. To force the 4 mA output, put 

its switch in the up position. Check the current output at 

the Power terminals, and if necessary trim to 4.00 mA using 

the  appropriate  trimpot.  Return  the  switch  to  the  down 

position, and repeat the process with the 20 mA switch.

Power

Sensor

4-20 mA

UP
DOWN

L

R

FTB700-T

Frequency

0

98

7 6 5 4

3

2

1

0

98

7 6 5 4

3

2

1

0

98

7 6 5 4

3

2

1

0

98

7 6 5 4

3

2

1

Power

Sensor

4-20 mA

4mA Adjust

Force 4 mA

Force 20 mA

20 mA Adjust

S

FTB700-T

Frequency

0

9

8

7 6 5 4

3

2

1

0

9

8

7 6 5 4

3

2

1

0

9

8

7 6 5 4

3

2

1

0

9

8

7 6 5 4

3

2

1

Setting Frequency Example

1. 

In an installation with an estimated maximum 

flow rate of about 150 GPM, a flow rate of 170 

GPM is selected as the full-scale maximum, the 

flow at which the current loop will register 20 

mA.

2.  In this example, the K-factor (found on the 

meter or fitting, or in the manual) is  “K = 

54.50”.      

3. 

Calculate the frequency as:

54.50 x 170 = 154.42
              60

4.  Rounding to one decimal point, enter 154.4 on 

the rotary switches by turning the rotary switch 

pointers to the desired digits.

Power

Sensor

4-20 mA

FTB700-T

Frequency

0

98

7 6 5 4

3

2

1

0

98

7 6 5 4

3

2

1

0

98

7 6 5 4

3

2

1

0

98

7 6 5 4

3

2

1

1

5

4 . 4

Switch Position

Seconds

L

R

2

down

down

4

down

up

8

up

down

16

up

up

Summary of Contents for FTB700

Page 1: ...e mail info omega com For latest product manuals www omegamanual info User sGuide Shop online at omega comSM PVC Stainless Steel FTB700 SERIES Inline Turbine Meters ...

Page 2: ...eering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide omega com info omega com The information contained in this document is believed to be correct but OMEGA accepts no liability for any errors it contains and reserves the right to alter specifications without notice ...

Page 3: ...pair Installation Page 5 Connections Page 5 Maintenance and Repair Page 5 Replacement Parts Page 6 FTB700 T BLIND ANALOG TRANSMITTER General Information General Information Page 7 Specifications Page 7 Features Page 8 Installation Installation Page 8 Connections Page 8 Settings and Calibration Setting Frequency Page 9 Setting Averaging Time Page 9 Checking Calibration Page 9 Troubleshooting Proble...

Page 4: ...e compatible for use with other remote mount Omega displays and controls as well GENERAL INFORMATION Specifications Electronic Options Specifications FTB700 FTB700 DPF144 P DPF144 W Power 6 36 Vdc 7 30 Vdc 7 30 Vdc 4mA 4 20 mA when loop powered Electronic Display DPF143 DPF144 Pulse Out 0 160 pulse sec current sinking Pulse scaled and pulse pass through Pulse scaled Analog Out 4 20 mA Loop Rate 5 ...

Page 5: ...EPAIR PVC Meter Stainless Steel Meter Maintenance and Repair Recalibration If it is necessary to recalibrate the meter for any reason please contact Omega Turbine Insert Removal and Installation CAUTION First remove all pressure from the line Then remove the screws or bolts that hold the insert in place and tug gently until the insert comes free A twisting motion can help to loosen the O ring seal...

Page 6: ...298 100298 10 Pickup 100419 100419 11a Insert SS N A 100440 11b Insert 3 PVC 100438 N A 11c Insert 4 6 PVC 100439 N A 12 Shaft assembly ceramic includes 2 103312 103312 13 Rotor PVDF bearing assembly 100333 100333 14 O ring EPDM 100214 100270 15 Insert screw 4 req d 100023 N A ALL METERS White Housing Blue Housing 1 Upper housing electronics Contact service representative for your specific model C...

Page 7: ...g is cast from aluminum and fuse coated The clamshell housing is provided with mounting feet or remote mounting The FTB700 T will operate on a relatively wide range of current loop voltages 24 to 36 Vdc Lower voltages limit the load that can be applied to the loop without distortion of the signal See Load Supply chart if there is a question regarding voltage vs load A built in power regulator supp...

Page 8: ...terminal blocks in the upper half which contains the potted electronics Installation Connections 0 9 8 7 6 5 4 3 2 1 0 9 8 7 6 5 4 3 2 1 0 9 8 7 6 5 4 3 2 1 0 9 8 7 6 5 4 3 2 1 Frequency Power Sensor 4 20 mA 4 20 mA Device e g Pump PLC Chart Recorder 24 36 Vdc Power Supply may be included in control unit Red White Black S FTB700 Series FTB700 T 4mA Adjust Force 4 mA Force 20 mA 20 mA Adjust FTB700...

Page 9: ...d their corresponding times Checking Calibration Normally it should not be necessary to check calibration since the digital design of this unit virtually eliminates drift However there are two types of calibration check that can be performed Look at the diagram below to locate the 4 and 20 mA force switches To force the 4 mA output put its switch in the up position Check the current output at the ...

Page 10: ...ty Check polarity Output stick at 4 mA No frequency input from flow sensor Check flow sensor connections Check flow sensor polarity Be sure terminal blocks are firmly plugged in With flow sensor disconnected use short wire to repeatedly short between sensor S and terminals Output should rise mA signal does not match flow rate Inadequate voltage Check load vs supply chart Wrong frequency setting Re...

Page 11: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Page 12: ...elocity Indicators M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Communications Based Acquisition Systems M U Data Logging Systems M U Wireless Sensors Transmitters Receivers M U Signal Con...

Reviews: