FTB700-T INSTRUCTIONS
Page 8
omega.com
GENERAL INFORMATION/INSTALLATION
Features
4-20 mA adjustment switches
Averaging time switches
Loop power indicator light
Potted for moisture protection
Easy to use rotary switches
for
frequency setting
Fusion coated cast aluminum housing
The upper portion of the housing must be removed to
make connections. Use a standard hex wrench (5/32” or 4
mm) to loosen the screws, then remove the upper half. The
connections are made to terminal blocks in the upper half,
which contains the potted electronics.
Installation
Connections
0
9
8
7 6 5 4
3
2
1
0
9
8
7 6 5 4
3
2
1
0
9
8
7 6 5 4
3
2
1
0
9
8
7 6 5 4
3
2
1
Frequency
Power
Sensor
4-20 mA
-
+
4-20 mA Device
(e.g. Pump, PLC,
Chart Recorder)
24-36 Vdc
Power Supply
(may be
included in
control unit)
Red
White
Black
-
+
-
+
S
FTB700-Series
FTB700-T
4mA Adjust
Force 4 mA
Force 20 mA
20 mA Adjust
FTB700-T Connection Diagram
Consult the FTB700-T Connections diagram before
connecting to the current loop and flow sensor. Be careful
to follow the color coding of the flow sensor wires in order
to establish the correct polarity. Incorrect polarity can
damage the sensor.