FMG490 INSTRUCTIONS
Page 14
CONNECTIONS
Remote Meter Wiring
1. On the remote meter, unscrew the user access lid
and remove it.
2. Remove the 5 pin screw connector from its socket.
3. Remove the plug and o-ring from the cable port.
4. Install cable gland and insert cable end.
5. Strip cable jacket and conductors and install
the wires into the 5 pin screw connector in their
respective locations (+Rem Pwr (red), - Rem Pwr
(black), B (orange), and A (blue)).
6. Secure the remote sensor cable inside the internal
strain relief clip and tighten the cable gland sealing
nut securely (torque strain relief/sealing gland
dome nut to 22 in-lbs.). A lose nut could cause
moisture ingress and compromise the meter head’s
IP68 rating; voiding the warranty. Plug the 5 pin
screw connector into its socket.
7. Reinstall the user access lid. Be sure to avoid cross-
threading the lid and to not pinch any wires with
the lid.
RED
BLACK
ORANGE
BLUE
Remote meter
15 pin connector in remote display
Remote Meter
Orange (R) Remote B: pn4
Blue (R) Remote A: pn5
R
Black (R) -Remote Pwr: pn2
Red (R) +Remote Pwr: pn1
User access lid
Front panel—
do not open
Cable Installation