Connect output signals and power
to this 4-terminal block.
2.
4.
3.
3.
Sensor Type
Pipe Size
Jumper Position
½ in. to 2½ in.
FMG3001
DN15 to DN65
FMG3002
3 in. to 4 in.
DN80
to
DN100
FMG3101
5 in. to 6 in.
FMG3102 DN125
to
DN150
8
in.
DN200
FMG3201
10 in. to 12 in.
FMG3202
DN250 to DN300
1.
Omega FMG-3000 Series Magmeter
1. Quick Start Guide
This manual contains the general installation, wiring and calibration data for the Omega FMG-3000 Series Magmeter with Frequency or
Current output. The basic steps are outlined on this page. See each referenced section for detailed information.
1.
Position the PIPE SIZE Jumper according
to your pipe size. Sec. 5 Pg. 4
2.
Install the Magmeter into the pipe.
Use Omega installation fi ttings ONLY.
The installation fi tting is critical to
Magmeter performance. Sec. 3-4, Pg. 3
3.
Connect
POWER
and
OUTPUT
wiring.
FMG-3001 and FMG-3101 Frequency out:
Sec. 8.1 Pg. 6
FMG3002 and FMG-3102 Current out wiring:
Sec. 7 Pg. 5
GROUNDING
Without a good Earth Ground, the Magmeter
may not operate effi ciently. Sec. 6 Pg. 5
4.
Route the wiring out through the two
cable ports. Use appropriate hardware
to secure the FMG-3000 from moisture
intrusion. One Liquid Tight Connector
is included. Sec. 5-6 Pg 4-5
Topic:
Page
1. Quick Start Guide
1
2. Specifi cations
2
3. Installation Pipe Fittings
3
4. Selecting a Location
3
5. FMG-3000 Magmeter Confi guration
4
Topic:
Page
6. General
Installation
&
Grounding
Tips
5
7. Wiring the Magmeter with 4-20 Loop out
5
8. Wiring the Magmeter with freq. out
6
9. Calibration Data
6-10
10. Maintenance & Troubleshooting
11
11. Ordering Information