background image

5

3. Electronic Evaluation and Electrical Safety

 

 

 

Does the 

supply voltage

 of the flow meter match the available power supply?

  

 

Is the supply voltage to the power supply adapter or the evaluating device sufficiently 

filtered

?

  

 

Does the 

output

 of the supply voltage correspond with the required output?

  

 

Is the electrical connection established based on the enclosed 

wiring plan 

(see page 19)?

  

 

Is the 

cable shield

 correctly grounded on both sides to a clean common ground (PE)?

  

 

Is there a 

potential difference

 between the ground on the flow meter and at the evaluating device?

  

 

Is the flow meter permanently 

grounded (PE)

 (e.g. via the pipelines)?

  

 

If the measuring element of the flow meter is 

insulated 

from ground (PE),  the meter must be grounded with a cable!!

  

 

Is the 

4-pin to 5-pin round pin plug 

of the connection cable firmly attached to the plug of the flow meter?

  

 

Are the wires at the 

evaluating

 device connected correctly?

  

 

Does the entire system meet the legal rules and regulations concerning electromagnetic compatibility 

(EMC)

?

  

 

Is compliance with all local rule and regulations, 

applicable rules

, guidelines and basic conditions of the 

EMC

 ensured?

  

 

Systems where a malfunction or failure may lead to personal injuries must be equipped with 

suitable safety mechanisms

. The function of these  

 

 

safety mechanisms must be checked at regular intervals.

6. Maximum Operating Pressure

Before installing the flow meter, you must check whether the max. operating 
pressure of the system does not exceed the max. permissible operating 

pressure of 450 bar of the flow meter. Make sure to keep in mind that peak 
pressures may occur when operating the system.

Important:

Please contact Omega with all operating pressures > 450 bar and in case of special models.  

7. Information about the EU Pressure Equipment Directive 97/23/EC 

Omega flow meters of the ”FHG“ series qualify as “pressure equipment“ as 
defined by Section 1, Paragraph 2.1.4. of the directive listed above and as 
such are affected by the regulations of this directive. 

Omega flow meters must therefore meet the technical requirements speci-
fied in Section 3, Paragraph 1.4 of the directive. The fluids to be measured 
are for the most part Group 2 fluids acc. to Section 9, Paragraph 2.2. 
Omega flow meters do not reach the limit values specified by Section 3, 
Paragraph 1.1. The technical requirements for Omega flow meters are    

therefore confined to the criteria specified in Section 3, Paragraph 3. This 
means that the devices must be designed and manufactured in accordance 
with the provisions of good engineering practice applicable in a member 
state. This is hereby confirmed. The section also stipulates that such pres-
sure equipment and components or accessories are not allowed to bear 
the CE marking in accordance with the Pressure Equipment Directive. This 
means that a declaration of conformity is not issued for Omega flow meters 
and the devices are not provided with the CE mark as pertaining to Direc-
tive 97/23/EC.

Important:

Verify that the specified maximum permissible operating pressure of the flow meter 

can never be exceeded in any operating mode of the system. Also pay attention to the 

flow measuring range, which is dependent on the viscosity of the fluid to be measured.

8. Flow Rate Measuring Range

The

 flow rate 

measuring range specified in the data sheet (Q

min

 – Q

max

of the flow meter refers to the test fluid ‚hydraulic oil‘ with a viscosity of 
21 mm

2

/s at a temperature of 20°C. For this measuring range, Omega 

specifies accuracy up to 0.3% of the measured value and a repeatability 
of 0.05%. 

In fluids with low viscosity (< 21 mm²/s), the measurement accuracy de-
grades while it may improve with fluid with a high viscosity (> 21 mm²/s). 
Note also that the flow measuring range is limited at higher viscosity (see 
data sheet of the flow meter). The characteristic pressure loss curves are 
listed in Section 23.

Summary of Contents for FHG Series

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info FHG FLOW METERS Shop online at User sGuide...

Page 2: ...ged by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free 0 800...

Page 3: ...OPERATING INSTRUCTIONS For FHG Series Flow Meters...

Page 4: ...ics 11 15 Signaling LEDs 12 16 Operating Mode Messages 12 17 Alarm and Warning Messages 13 18 Preamplifier Technical Data 13 19 Preamplifier Pin Assignment 13 20 Maintenance 14 21 Returning for Repair...

Page 5: ...ply them by a high resolution interpolator using adjustable settings The square wave signals are bidirectional and can be utilised by any evaluating instrument as well as computers and PLC controls Th...

Page 6: ...tion check and verify the following properties aspects of the respective circumstances of your system to ensure operation is trouble free safe and reliable 1 The Fluid to be Processed Are the flow met...

Page 7: ...qualify as pressure equipment as defined by Section 1 Paragraph 2 1 4 of the directive listed above and as such are affected by the regulations of this directive Omega flow meters must therefore meet...

Page 8: ...Thread Please comply with the screw in depths and sealing systems PTFE tape or liquid sealants such as adhesives are not permitted Fastening The devices must be installed stress free into the pipeline...

Page 9: ...ply them by a high resolution interpolator using adjustable settings The square wave signals phase shif ted by 90 are bidirectional and can be utilised by any evaluating device as well as computers an...

Page 10: ...surement volume Vm cm3 Imp K Factor Imp l K Factor Imp gal 1 0 10 00000 100 379 2 1 5 00000 200 757 5 2 2 00000 500 1893 10 3 1 00000 1000 3785 25 4 0 40000 2500 9464 32 5 0 31200 3200 12113 50 6 0 20...

Page 11: ...9 FHG 1xx2 FLOW vs FREQUENCY FHG 1xx4 FLOW vs FREQUENCY Flow diagrams vs frequency...

Page 12: ...xample Flow meter FHG 1xx4 Max processable input frequency of the downstream evaluating unit 20 kHz Max operating flow 140 l min Path 1 The diagram yields an IPf of 25 Path 2 IPF fmax x VmIPF1 x 60 Qm...

Page 13: ...ser is able to set the degree of filtering in the form of partial volumes using rotary coding switches Adjustable pulse filtering 0Z 0 25Z 0 5Z 0 75Z 1 0Z 1 25Z 1 5Z 1 75Z 2 0Z 2 25Z 2 5Z 2 75 3 0Z 3...

Page 14: ...fly press the S2 key for the acknowledgment The new pulse rate is enabled a once The rotary coding switch for the pulse filtering has 16 switch settings The degree of filtering is determined with quar...

Page 15: ...tive flow 6 Temperature errors The temperature of the fluid is too high 120 C and may result in flawed or incorrect measurements 18 Preamplifier Technical Data Scanning sensor 2 x GMR sensors in a bri...

Page 16: ...ly well shielded connection cables with a wire cross section of 4 to 5 x 0 25 mm Please note that the housing of the round pin plug is metallic has a connection for the shield and that the potential o...

Page 17: ...flow meter has been completely cleaned and flushed This is to protect our employees and makes our work easier In case of noncompliance with this rule the devices are returned to the sender without att...

Page 18: ...sswiderstand p Durchflusswiderstand p Durchflussbereich 0 bis 50 l min Flow range 0 up to 120 l min Flow range Q Flow range Q Flow range Q Flow range Q Flow range Q Flow range pressure drop p Flow ran...

Page 19: ...sseinheit AR 2500 R Flanschanschlussma e DIN EN 1092 Gewicht 120 kg FHG1xx5 FHG1xx7 Preamplifier Preamplifier Earthing Earthing Weight 81 kg Weight 81 kg Weight 120 kg Sensor module Sensor module Conn...

Page 20: ...ier Preamplifier Earthing Earthing on both sides on both sides M8 15 deep M8 15 deep Weight 12 kg Weight 12 7 kg Weight 24 8 kg Weight 22 kg Sensor module Test Port Test Port Sensor module Connection...

Page 21: ...2 Evaluating unit e g display Flow sensor FHG Q 8l min V 11 53 l direction pos f 424 9 Hz control input 0 voltage blue channel 1 white channel 2 black Error Direction grey 25 Pin Assignment Channel 1...

Page 22: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 23: ...city Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industria...

Reviews: