background image

PRODUCT INSTRUCTION SHEET

  
Operating temperature range:    

0-45 degC

  Maximum operating pressure:    

1 bar/14.7  psi/1atm

  Flow rate minimum:   

 

.20 gpm

  Flow rate maximum:  

 

0.6 ppm

  pH range:  

 

 

5.5-9.0

  Output signal: 

 

 

4.0+/- 0.2mA in air (zero)

 

 

 

 

20mA +/- 0.2mA at high range

     

(2, 

or 

10ppm)

  Power Requirement:  

 

12-24 VDC, 

250

 mA minimum

  Cross-Sensitivity: 

  

 

ClO2, ozone, bromine 

  Chemical Compatibility: 

 

up to 50% ethanol/water or

    up 

to 

50% 

glycerol/water 

 

 

    

 OPERATING SPECIFICATIONS

SECTION 12.0

SENSOR SPECIFICATION

12.1 OPERATING SPECIFICATIONS.   Follow all operating specifications, 
especially for pH and flow rate as noted in the specification tables 
below.

  
   SENSOR
    Dimensions:   

 

 

8.2"L x 1" dia

    Body Material:  

 

 

Black PVC

    Membrane Material:  

 

polyethersulfone

    O-ring material: 

 

 

Viton

R

    Cathode: 

 

 

Gold

    Anode:    

 

 

Silver chloride (AgCl)

    Cable: 

 

 

 

2 -conductor shielded, 10ft (3mtr)

     

tinned 

wire 

leads

  FLOW CELL
    Dimensions:   

 

 

5.58"H x 2.25"DIA

    Material:  

 

 

Acrylic

    Connections: 

 

 

1/4" NPT inlet and outlet

 

 

    TECHNICAL SPECIFICATIONS

 

         MAINTENANCE/REPLACEMENT PARTS

   FC72 

Flow Cell, 1/4 inch FNPT inlet and outlet, includes:  

                                       2 each 1/4" barbed tube fittings, clamp, threaded  
 

flow cell installation fitting

   FCL-ES   

Free Chlorine sensor fill solution, 30mL, 1 each

   FCL-RM   

FCL Replacement premembraned cap, 1 each 

FC72 FLOW CELL DIMENSIONS

SENSOR AND FLOW CELL INSTALLATION DIMENSIONS

Parts covered by this product instruction sheet include: FCLTX-100 Series

 M4679/0708 page 6 of 6