background image

PRODUCT INSTRUCTION SHEET

10.1 GENERAL MAINTENANCE.  Service the sensor regularly 
(weekly is suggested)

10.2 MEMBRANE CAP REPLACEMENT.  If membrane replace-
ment is required, a new cap with preinstalled membrane must 
be used.  Order FCLA-2016 replacement membrane cap and 
FCLA-2015 refill solution.  To change membrane cap (do the 
following over a sink or washbasin):  
a) turn sensor upside down with cap facing upward
b) rotate cap counter-clockwise to remove (see FIG 10).  
c) fill sensor body with electrolyte using needle and bottle of 
refill solution (see FIG. 11)
d) install a new membrane cap by threading cap onto sensor 
rotating cap clockwise (opposite of FIG 10)

10.3 MEMBRANE CAP/SENSOR CLEANING.  Rinse cap with 
water only.  If cap does not clean, replace with new one.

SECTION 10.0

SENSOR MAINTENANCE/RECONDITIONING

SECTION 11.0

SENSOR TROUBLESHOOTING

11.1 CALIBRATION PROBLEMS
 

a) Sensor output 

HIGHER

 than DPD test

 

 

1) Run in time too short

 

 

2) Membrane cap damaged

 

 

3) Interference from water 

 

 

     contaminants (see Specifications,

 

 

      "Cross Sensitivity")

 

 

4) Cable short circuit or damage

 

 

5) pH value less than pH 5.5

 

b) Sensor output 

LOWER

 than DPD test

 

 

1) Run in time too short

 

 

2) Deposits on Membrane cap 

 

 

3) Flow rate too low

 

 

4) Air bubbles on membrane

 

 

5) Surfactants in water

 

 

6) pH value more than pH 8.0

 

 

7) No electrolyte in membrane cap

 

 

 

c) Sensor ouput is 4mA (zero ppm)

 

 

1) Run in time too short

 

 

2) Only bound chlorine present

 

 

3) Chlorine content below detection

 

 

      limit

 

 

4) Sensor not wired correctly (See 

 

 

     SECTION 6.0 of this manual)

  5) 

Defective 

sensor

 
 

d) Sensor output 

UNSTABLE

 

 

1) Air bubbles on membrane

  2) 

Membrane 

damage

  3) 

Non-sensor 

problem

 

  

Symptom

 

 

Possible Cause

 

 

 

Solution/Remedy

  The sensor   

1) Run in time too short   

1) See Sec 5.0 -CONDITIONING

  cannot be    

2) Membrane cap damaged 

2) Replace cap - See Sec 8.0

  calibrated-  

3) Interference from contaminants 

3) See SPECIFICATIONS

  ouput is  

 

4) DPD chemicals bad 

 

4) Use new DPD kit

  HIGHER than

 

5) pH value < pH 5.5 

 

5) Increase pH (5.5-8.0)   

  DPD Test   

 

 

 

     See SPECIFICATIONS

 

 

6)Temperatue increased since cal 

6) Match calibration temp.

  The sensor  

1) Run in time too short   

1) See Sec 7.0 -CONDITIONING

  cannot be   

2) Deposits on membrane cap 

2) Remove deposits or replace 

  calibrated-  

 

 

 

      cap if cleaning ineffective.

  output is    

3) Flow rate too low 

 

3) increase fl ow - See  

  LOWER than

 

 

 

 

    SPECIFICATIONS

  DPD Test   

4) Air bubbles on membrane 

4) Remove and re-install sen 

 

 

 

 

 

      sor to remove bubbles.

 

 

5) Surfactants in water   

5) Remove surfactants and 

 

 

 

 

 

      replace cap 

 

 

6) pH > pH 8.0 

 

6) Lower pH (5.5-8.0)  

 

 

 

 

 

    See SPECIFICATIONS

 

 

7) No electrolyte in cap   

7) Add new electrolyte, run in

 

 

 

 

      

     sensor and re-calibrate

 

 

8) Temperature decreased since cal  8) Increase temp to match cal

 

 

9) Organic chlorination agents 

9) Use chlorinating agents

 

 

    present in water 

                              per DIN 19643

  Sensor output 

1) Only bound chlorine present 

1) Check for Chloramine with 

  is 4mA (0 ppm) 

     NO FREE CHLORINE 

 

     appropriate DPD test.  

 

 

 

 

 

     Replace water/ Rechlorinate.

 

 

2) Run in time too short   

2) See Sec 7.0 -CONDITIONING

 

 

3) Chlorine content below limit 

3) Add chlorine and repeat  

 

 

 

 

     

      calibration

 

 

4) No electrolyte in cap   

4) Refi ll electrolyte

 

 

5) Sensor electrical connection 

5) See SECTION 6.0

 

 

     wrong

  Unstable output 

1) Air bubbles on sensor membrane  1) Tap to remove bubbles

  from sensor 

2) Membrane damaged   

2) Replace membrane, run in

 

 

 

 

 

     sensor and recalibrate.

 

 

3) Non-sensor problem   

3) check PLC or I/O device

 

 

  

TROUBLESHOOTING CHART

FIG. 10

 

INJE

C

T UNTIL 

S

OLUTION 

LEAK

S

 OUT OF FILL HOLE   

FIG. 11

Parts covered by this product instruction sheet include: FCLTX-100 Series

 M4679/0708 page 5 of 6

Reviews: