background image

Power Supply

PLC, DVM,...

sensor red wire

sensor 
black wire

FIG. 3

PRODUCT INSTRUCTION SHEET

6.0 ELECTRICAL  INSTALLATION.  The sensor produces an ap-
proximate output of 4 mA in air and 20mA at the top range of 
free chlorine output (0-2ppm,  0-5ppm, 0-10ppm).
  

NOTE: The supply voltage to the Sensor must be 12-24 V DC with 
minimum of 250 mA.  Maximum load is 1 Watt.  The sensor has 3 wires, 
red (+) , black (-) and clear  (shield).  Twist together or solder black 
and clear if  instrument does not have separate ground.  If a separate 
ground is available such as for PLC’s connect clear (shield) to it.
Attach the red wire to the power supply positive terminal (+)and the 
black wire to the PLC or DVM positive (+) terminal. Connect a wire 
(customer supplied) from the power suppy negative (-) and the PLC or 
DVM (-).  See FIG 3.   The Sensor will require several minutes to stabilize 
after power is supplied to it.

7.0 SENSOR CONDITIONING The sensor requires conditioning 
prior to generating stable values. 
 

a. For new Sensors, allow  the Sensor to run for at least

 

4 hours before calibration.

 

b. If the Sensor will be un-powered for 2 hours or  

 

 

more, run for 3 hours prior to use.

 

c. After membrane/electrolyte replacement, allow the  

 

Sensor to run for at least 4 hours.

SECTION 6.0

ELECTRICAL INSTALLATION

SECTION 7.0

SENSOR CONDITIONING

Parts covered by this product instruction sheet include: FCLTX-100 Series

 M4679/0708 page 3 of 6

Reviews: