SECTION 3.0

SENSOR PREPARATION

FIG. 2

3.0 FREE CHLORINE SENSOR ASSEMBLY.  The Free Chlorine  
Sensor is shipped with the membrane cap pre-installed and 
covered with a cap with water inside to keep the membrane 
wet.  Make sure to keep sensor cap, anode and cathode inside 
the sensor body, away oily or greasy materials.  Contact with oil 
or grease will result in inaccurate measurements.

NOTE: IF SENSOR WILL BE STORED DRY OUT OF FLOW  

 

CELL, SHAKE BODY DOWNWARD INTO A SINK TO REMOVE THE 
FILL SOLUTION.  TAKE THE MEMBRANE CAP AND IMMERSE IN A 
CUP OF TAP WATER UNTILL READY TO REUSE.   SEE    
SECTION 9.  REPLACE CAP AND ELECTROLYTE BEFORE INSTALL-
ING INTO FLOW CELL (see SECTION 10 for cap and electrolyte 
change.  See SECTION 5 for sensor installation into flow cell).

SECTION 4.0

FLOW CELL INSTALLATION

4.0 FLOW CELL. To obtain accurate Free Chlorine reading, 
the Sensor must be installed into the Flow Cell to prevent air 
bubbles formation on the membrane, proper spacing between 
the sensor and the installation wall, and laminar flow across 
the membrane.

4.1. Using two 1/4” NPT Tube fittings, connect the FC72 Flow             
        Cell into your system, noting the inlet (bottom) and outlet   
        (side) orientation.  (see FIGURE 2)
4.2.  Install clamp with rubber backing as shown in FIG. 2A.
4.3  Drill 3/8" diameter hole on the panel.
4.4  Insert bolt as shown in FIG 2A
4.5  On back of panel attach lock washer and nut to secure
        clamp and flow cell to panel. 

PRODUCT INSTRUCTION SHEET

SECTION 5.0

SENSOR INSTALLATION

5.0 SENSOR INSTALLATION INTO FLOW CELL.   
 

a. First install threaded fitting onto sensor body  

 

 

(remove fitting if pre-installed in flow cell)  

 

b.  Install snap-ring into groove on sensor body

 

c. Next, slide o-ring onto body of sensor until it  

 

 

reaches bottom of threaded fitting.

 

d. Thread sensor assembly into top of flow cell as 

 

     shown in FIGURE 2.

 

d. Turn on flow and verify the flow through the Flow  

 

    Cell is at least 0.2gpm (45 liters/hour and no

 

     more than 0.6gpm (135 liters/hour).

 

 

TH

R

EADED FITTING

O-

R

ING

S

NA

P

 

R

ING (IN

S

TALL

S

 

INTO 

PR

OBE G

R

OOVE)

S

NA

P

 

R

ING 

G

R

OOVE

BA

R

BED TUBE 

      FITTING

BA

R

BED TUBE 

      FITTING

FLOW 

C

ELL

S

EN

S

O

R

FIG. 2A

FC72 Flow cell

Parts covered by this product instruction sheet include: FCLTX-100 Series

 M4679/0708 page 2 of 6

Reviews: