![background image](http://html2.mh-extra.com/html/omega/dp26-rtd/dp26-rtd_user-manual_4044223059.webp)
ANALOG OUTPUT (if applicable)
Signal type:
Current or voltage
Signal level:
Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function:
May be assigned to a display range or proportional
control output with setpoint #1 when used as a
control output.
Linearity:
0.2%
Load Regulation:
1.1%
Line Regulation:
0.02% /V ac
Step Response Time:
2 - 3 seconds to 99% of the final value
ALARM OUTPUTS (if applicable)
2 Form "C" on/off relays. Configurable for latched and unlatched by software.
Max current:
5 AMPS, resistive load.
Max voltage:
250 V ac or 30 V dc
ISOLATED ANALOG OUTPUT (TB5, if applicable)
Same as non-isolated analog output except isolated to 1000 Vdc.
Signal Type:
Current or voltage
Signal Level:
Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function:
May be assigned to a display range or proportional
control output with setpoint #1 when used as a
control output.
Linearity:
0.2%
Step Response Time:
2 - 3 seconds to 99% of the final value
Only one analog output is available on each unit and it must be factory installed.
51
Summary of Contents for DP26-RTD
Page 4: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......
Page 52: ......
Page 53: ......
Page 54: ......