OPERATION   

 

Preparation 
 

1.

 

The heater design, temperature sensor and control circuit give good temperature control and 
uniformity, but make sure that there is a close fit of the probes in the block to allow efficient heat 
transfer. Contact us about an insert that more closely fits your probe or device being calibrated. 

 

2.

 

Plug the power cable into the socket in the back of the unit. Connect the power cable to the 
electrical supply and switch the power on. 1 = power on, 0 = power off. 

 

3.

 

Clean the heater block cavity out with 
shop or canned air to remove any 
particulate. Next place the probe insert 
into the heater block as shown using the 
supplied insert extractor to minimize the 
risk of damaging the heater block and/or 
probe insert. Never place a hot insert into 
a cold heater block or vice versa as the 
insert may become jammed which will 
damage both parts. Always use the insert 
extractor to both install and remove the 
probe insert. 

 

4.

 

To prevent damage to the heater block, 
insert, heaters and PRT block sensor DO 
NOT use the following in or around the 
block; 
Oil, Thermal grease, Water 
Aluminum oxide sand,

 

Ceramic fiber insulation or Kaowool

 

 

 

 
Setting the operating temperature 

1.

 

To set the operating temperature required, press and hold either the up or down arrow button to 
increment to the value required. The values will increment faster as you hold down the button.  

2.

 

When you have the correct set temperature displayed the unit will start to heat or cool to that 
value. 

3.

 

Once the process value/actual temperature reaches the set point, allow the block to fully stabilize 
for at least 15 minutes before performing a calibration. 

4.

 

Upon completion of your work set the temperature to 50°C/122°F or less and allow it to cool before 
transporting or moving. The block fan will kick on to provide cooling. After a safe temperature has 
been reached power can be switched off and the unit unplugged.  

 

 

 

 
 

 

 

Summary of Contents for DBCL400

Page 1: ...il info omega com For latest product manuals www omega com Above items are included with the DBCL400 Dry Block temperature calibrator Not shown is the calibration certificate and operator manual Model...

Page 2: ...gnetic interference and electrical safety regulations Specification Figures quoted are at the base of the well at the time of calibration Temperature range 5 C 9 F above ambient to 450 C 850 F Over te...

Page 3: ...safety problem is encountered switch off at the power socket and remove the plug from the supply Please use caution when removing probes and inserts as burns to the skin can occur if in contact Insta...

Page 4: ...mage both parts Always use the insert extractor to both install and remove the probe insert 4 To prevent damage to the heater block insert heaters and PRT block sensor DO NOT use the following in or a...

Page 5: ...HIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL REMOVING THE FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL VOLTAGES THERE ARE NO OPERATOR MAINTAINABLE PARTS WITHIN THE EQUIPMEN...

Page 6: ...oles DBCL 400 3044 Insert 2 x 1 4 2 x 3 8 holes DBCL 400 3049 Insert 1 x 5 8 1 x 1 4 holes DBCL 400 3045 Insert 2 x 1 4 2 x 1 2 holes DBCL 400 3050 Insert 1 x 11 16 1 x 1 4 holes DBCL 400 3046 Insert...

Page 7: ...U The Electromagnetic Compatibility Directive 2011 65 EU The Restriction of Hazardous Substances Directive 6 References to the relevant harmonised standards used or references to the other technical s...

Page 8: ...78 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in...

Page 9: ...y OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limite...

Page 10: ...s M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U...

Reviews: