background image

5 | 

P a g e

 

 

Temperature scale conversion  

To change the temperature scale press        to display the “UNIT” parameter and change as needed. Next 
change the remaining controller values as shown in the table below based on degrees C or F.  

Unit serial number = 

 

Parameter 

Operation in Degrees C 

Operation in Degrees F 

CALO 

50 

122 

CAHI 

400 

752 

OFTL 

 

 

OFTH 

 

 

 

Calibration

 

The unit has been calibrated by the factory to meet specifications. In the event that you want to adjust or 
correct the calibration use the following parameters with the display unlocked. 
Press        and OFTL will display which is the Zero or low end adjustment. Enter a negative value to correct for 
low readings and vice versa. For example if your reference thermometer is showing that the DBCL400 is 2.0 

degrees low then enter a -2.0. Press 

  to access OFTH which is the span or high end correction. Use a 

negative value for readings that are low.  
 

Operator maintenance 
 

NOTE  THAT  THIS  EQUIPMENT  SHOULD  ONLY  BE  DISMANTLED  BY  PROPERLY  TRAINED  PERSONNEL. 
REMOVING  THE  FRONT  OR  REAR  PANELS  EXPOSES  POTENTIALLY  LETHAL  VOLTAGES.    THERE  ARE  NO 
OPERATOR MAINTAINABLE PARTS WITHIN THE EQUIPMENT

.

  

 

 

In  the  unlikely  event  that  you  experience  any  problems  with  your  unit  which  cannot  easily  be  remedied,  you 
should  contact  your  supplier  and  return  the  unit  if  necessary.  Please  include  any  details  of  the  fault  observed 
and remember to return the unit in its original packing. Omega Engineering will accept no responsibility for any 
damage to units that are improperly packed for shipment.  If in doubt, contact your supplier.  
 

1.

 

Cleaning:  Before cleaning your unit, ALWAYS disconnect it from the power supply and allow it to 
cool below 50° C. Your unit can be cleaned by wiping with a damp soapy cloth. Care should be 
exercised to prevent water from running inside the unit. Do not use abrasive cleaners. 

2.

 

Fuse:  The unit is protected by a fuse. It should only be changed by suitably qualified  
personnel. If the fuse blows persistently, a serious fault is indicated and you may need to return the 
unit to your supplier for repair. 

 

 
 
 
 

Summary of Contents for DBCL400

Page 1: ...il info omega com For latest product manuals www omega com Above items are included with the DBCL400 Dry Block temperature calibrator Not shown is the calibration certificate and operator manual Model...

Page 2: ...gnetic interference and electrical safety regulations Specification Figures quoted are at the base of the well at the time of calibration Temperature range 5 C 9 F above ambient to 450 C 850 F Over te...

Page 3: ...safety problem is encountered switch off at the power socket and remove the plug from the supply Please use caution when removing probes and inserts as burns to the skin can occur if in contact Insta...

Page 4: ...mage both parts Always use the insert extractor to both install and remove the probe insert 4 To prevent damage to the heater block insert heaters and PRT block sensor DO NOT use the following in or a...

Page 5: ...HIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL REMOVING THE FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL VOLTAGES THERE ARE NO OPERATOR MAINTAINABLE PARTS WITHIN THE EQUIPMEN...

Page 6: ...oles DBCL 400 3044 Insert 2 x 1 4 2 x 3 8 holes DBCL 400 3049 Insert 1 x 5 8 1 x 1 4 holes DBCL 400 3045 Insert 2 x 1 4 2 x 1 2 holes DBCL 400 3050 Insert 1 x 11 16 1 x 1 4 holes DBCL 400 3046 Insert...

Page 7: ...U The Electromagnetic Compatibility Directive 2011 65 EU The Restriction of Hazardous Substances Directive 6 References to the relevant harmonised standards used or references to the other technical s...

Page 8: ...78 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in...

Page 9: ...y OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limite...

Page 10: ...s M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U...

Reviews: