Omega Model CYD211 User’s Manual
Remote Operation
4-19
DFLT
Factory Defaults Command
Input:
DFLT 99[term]
Remarks:
Sets all configuration values to factory defaults and resets
the instrument. The "99" is included to prevent accidentally
setting the unit to defaults.
DISPFLD
Displayed Field Command
Input:
DISPFLD <source>[term]
Format:
n
<source> Specifies input data to display. Valid entries:
0 = Kelvin, 1 = Celsius, 2 = sensor units, 3 =
Fahrenheit.
Example: DISPFLD
1[term]
– Displays Kelvin reading for the input.
DISPFLD?
Displayed Field Query
Input:
DISPFLD?[term]
Returned:
<source>[term]
Format:
n
(Refer to command for description)
FRDG?
Fahrenheit Reading Query
Input:
FRDG?[term]
Returned:
<temp
value>[term]
Format:
±nnnnnn
Remarks:
Also see the RDGST? command.
INCRV
Input Curve Number Command
Input:
INCRV <curve number>[term]
Format:
nn
<curve number> Specifies which curve the input uses. If
specified curve parameters do not match
the input, the curve number defaults to 0.
Valid entries: 0 = none, 1
–
20 = standard
curves, 21 = user curve.
Remarks:
Specifies curve the input uses for temperature conversion.
Example: INCRV
21[term]
– The input User Curve 21 for
temperature conversion.