Omega Model CYD211 User’s Manual
4-20 Remote
Operation
INCRV?
Input Curve Number Query
Input:
INCRV?[term]
Returned:
<curve number>[term]
Format:
nn
(Refer to command for description)
INTYPE
Input Type Parameter Command
Input:
INTYPE <sensor type>[term]
Format:
n
<sensor type> Specifies input sensor type. Valid entries:
0 = Silicon Diode
3 = 100
Ω
Platinum/500
1 = GaAlAs Diode
4 = 1000
Ω
Platinum
2 = 100
Ω
Platinum/250 5 = NTC RTD
Example: INTYPE
0[term]
– Sets input sensor type to silicon diode.
INTYPE?
Input Type Parameter Query
Input:
INTYPE?[term]
Returned:
<sensor
type>[term]
Format:
n
(Refer to command for description)
KEYST?
Keypad Status Query
Input: KEYST?
[term]
Returned:
<keypad
status>[term]
Format:
n
1 = key pressed, 0 = no key pressed.
Remarks:
Returns keypad status since the last KEYST?. KEYST?
returns 1 after initial power-up.
KRDG?
Kelvin Reading Query
Input:
KRDG?[term]
Returned:
<Kelvin value>[term]
Format:
+nnnnnn
Remarks:
Also see the RDGST? command.