16
Code
Content
Range
FIgS
Conductivity start value
0~1999uS/cm (k=1.0)
FIgE
Conductivity end value
0~1999uS/cm (k=1.0)
FICS
Temperature start value
0~100
FICE
Temperature end value
0~100
4.3.4 SHSP—Measurement Mode
CDTX-111 has conductivity, resistivity, TDS and temperature four modes of
measurement and display functions, it can be selected in the SHSP program.
Select SHSP in the menu, press
ENTER
into it and display SH 0 or SH 1 or
SH 2. SH 0 is conductivity mode, SH 1 is resistivity mode, SH 2 is
temperature mode. Press
UP
to switch it, then Press
ENTER
to store
and return to the setting state. The indicator light also changes to indicate
that the display state changes. The corresponding parameters in the setting
also will automatically change.
*NOTE The sensor should be changed when switch the measurement mode.
Code
Content
Indicator light
Cell constant
SH 0
Display conductivity
value
uS(mS)/cm
lights
K=0.01/0.1/1.0
SH 1
Display resistivity value
M
cm lights
K=0.01
SH 2
Display Temperature
value
C/F lights
K=0.01/0.1/1.0
4.3.5 FACt—Restore Factory Setting
Select FACt in the menu, press
ENTER
into it and display HHHH. At this
moment the instrument is being restored factory setting, about 10 seconds
it will automatically returns to FACt, restore factory setting is completed.
After this process, all value the user set before becomes
the factory
Table 8
Measurement
Mode
Table 7
Set input range for the current output