3

 

 

  

Figure 3-1.  Low Impedance  Accelerometer

 

 

LOW IMPEDANCE MICRO  CIRCUIT

 

 

OMEGA  has  a  broad  line  of  micro circuits  which  yield  excellent  low 
impedance output  characteristics.  These  hybrid  circuits  often  include  special 
filtering  capabilities  and  protection  schemes.    The  specific  micro  chip 
properties are  engineered to  create  the  output  characteristics  desired.  They 
are  extremely  rugged  and  capable  of  withstanding  extremely  high  shock 
without  damage or  degradation  of  performance

.   

Additional circuit  protection 

schemes Include:

 

 

•  ESO  protection

 

•  cross wire  protection

 

•  over  voltage  protection

 

When  used  properly within  specified  limits,  OMEGA accelerometers will

 

provide years  of  reliable,  trouble-free service. 

 

 

NOISE IN  ACCELEROMETERS

 

Electrical Grounding

 

 

An  accelerometer  may  be  designed  with 
the  output  signal  ground either  connected 
to  the  case  of  the  accelerometer 

or 

isolated 

from 

the 

case 

of 

the 

accelerometer.

 

Case Grounded  Accelerometers

 

Case 

grounded 

accelerometers 

ar e 

connected electrically (see  Figure  3-2)

 

to  any  conductive  surface  on  which  they 
are  mounted.    When  these  units  are  used, 
care  must  be  exercised  to  avoid  errors 
due  t o   ground 

noise.    These  case 

grounded  accelerometer cases  should  not 
be  allowed  to  conduct  to  any  outside 
potential or  noise  source.

 

 

The  case  ground  type  accelerometer should  not  be  selected for  applications 

where  ground loops  might  be  problem, such  as  machinery  monitoring.  The 
OMEGA Model  ACC  103 is  a case  grounded accelerometer.

 

Figure 3-2  Case 
Grounded  Accelerometer

 

Summary of Contents for ACC101

Page 1: ...e mail info omega com For latest product manuals www omegamanual info Shop online at omega comSM User sGuide ACC101 ACC102A ACC103 ACC301A ACC310 ACC320 Accelerometers...

Page 2: ...rves the right to alter specifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Page 3: ...TS 5 Power Supplies 5 SECTION 5 CONNECTOR WIRING 7 Coaxial Connectors 7 Two Pin Connectors 7 Four Pin Connectors 7 SECTION 6 CABLING TECHNIQUES 8 Splicing and Extending Cables 8 Securing Cables 8 SECT...

Page 4: ...jj...

Page 5: ...pment inspect the container and equipment for any signs of damage Take particular note of any evidence of rough handling in transit Immediately report any damage to the shipping agent NOTE The carrier...

Page 6: ...r case The materials and physical construction used iIn an accelerometer design are selected to furnish the particular performance characteristics desired For more information about the construction o...

Page 7: ...will provide years of reliable trouble free service NOISE IN ACCELEROMETERS Electrical Grounding An accelerometer may be designed with the output signal ground either connected to the case of the acce...

Page 8: ...the accelerometer will transduce shock and vibration that is input normal to the accelerometer s mounting base It shouldbe noted that the accelerometer will have some sensitivity to motion in the othe...

Page 9: ...celerometer Is not over powered POWER SUPPLIES If the data acquisition system being used does not incorporate its own power supply then a separate power supply must be used OMEGA manufactures single c...

Page 10: ...ow Impedance Wiring Directly to Instrumentation For those applications where the data acquisition system does not supply the constant current power an external power supply must be used Essentially th...

Page 11: ...l connectors are wired The result Is that the two pins correspond to the center conductor and shield of a coaxial accelerometer The two pin configuration lends itself to be wired with either coaxial c...

Page 12: ...ypically low impedance devices are not affectedby sharp cable bends or cable motion Figure 6 1 Cable Routing and Securing SPLICING EXTENDING CABLES Common sense and care should be used when splicing a...

Page 13: ...or noise source Letting a connector which is signal common touch ground can cause ground loop noiseInterjection When working with permanently routed cable systems it is recommended that the splices be...

Page 14: ...have a mounting stud protruding from the bottom of the unit This stud may or may not be removable depending upon the model When attaching the accelerometer with the mounting stud you should make certa...

Page 15: ...s lend themselves to be adhesively bonded to an object Temporary bonds can be achieved with many different tacky adhesives beeswax or cyanoacrylate type cements such as Krazy Glue These types of bonds...

Page 16: ...ets are not generally recommended for measurements above 5KHz They are quite adequate for lower frequency measurements andcanbe helpfulwhen a single accelerometer is used to measure a great number of...

Page 17: ...13 Figure 7 3 Magnetic Mounting...

Page 18: ...g and 105mV g The actual sensitivity for each unit is verified at 100 Hz and recorded on the accelerometer s calibration certificate BASIC FREQUENCY RESPONSE The frequency response specification for...

Page 19: ...y 1 Temperature Range 10 to 200 F 20 to 95 C Transverse Sensitivity 5 Shock Limit 5 000 g pk Mounted Resonant Frequency 18 kHz Output Impedance 100 ohms Bias Voltage 12 VDC typ Base Strain 250 in in 0...

Page 20: ...0 to 200 F 20 to 95 C Transverse Sensitivity 5 Shock Limit 5 000 g pk Mounted Resonant Frequency 18 kHz Output Impedance 100 ohms Bias Voltage 12 VDC typ Base Strain 250 in in 001 g pk in in typ Noise...

Page 21: ...ity 2 Temperature Range 40 to 250 F 40 to 121 C Transverse Sensitivity 5 Shock Limit 5 000 g pk Mounted Resonant Frequency 50 kHz Output Impedance 1000 ohms Bias Voltage 7 VDC typ Base Strain 250 in i...

Page 22: ...e Range 40 to 250 F 40 to 121 C Transverse Sensitivity 5 Shock Limit 5 000 g pk Mounted Resonant Frequency 25 kHz typ Output Impedance 1000 ohms Bias Voltage 7 VDC typ Base Strain 250 in in 005 g pk i...

Page 23: ...ty 1 Temperature Range 40 to 250 F 40 to 121 C Transverse Sensitivity 5 Shock Limit 5 000 g pk Mounted Resonant Frequency 40 kHz Output Impedance 100 ohms Bias Voltage 12 VDC typ Base Strain 250 in in...

Page 24: ...Temperature Range 40 to 250 F 40 to 121 C Transverse Sensitivity 5 Shock Limit 5 000 g pk Mounted Resonant Frequency 18 kHz Output Impedance 100 ohms Bias Voltage 12 VDC typ Base Strain 250 in in 005...

Page 25: ...21 NOTES...

Page 26: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 27: ...ine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industria...

Reviews: