Specifications
7
(Specifications are for all models except where noted)
THERMOMETER
Measuring:
OS530HR,
-22°F to 250°F (-30°C to 121°C)
Temperature
OS531:
0°F to 750°F (–18°C to 400°C)
Range:
OS530L,
OS533,OS532: 0°F to 1000°F (–18°C to 538°C)
OS534
0°F to 1600°F (–18°C to 871°C)
OS523
0°F to 2500°F (–18°C to 1371°C)
OS524
1000°F to 4500°F (538°C to 2482°C)
OS531:
±2% of reading or
3°F whichever is greater
All others
±1% of reading or
3°F whichever is greater
(2% Rdg for temp > 2000°F for OS524)
Field of Vision:
OS531, OS532, OS530L
10:1
OS533, OS530HR
20:1
OS534
30:1
OS53x-CF
.15"@6"
OS523-1
30:1
OS523-2
60:1
OS523-3
68:1
OS524
110:1
Repeatability:
± (1% rdg + 1 digit)
Resolution:
1°F or 1°C (0.1°F or °C for OS530HR)
Response Time:
250 msec
Spectral Response:
8 to 14 microns (2 to 2.5 microns, OS524)
(5 to14 microns, OS530L, OS531, OS532)
Thermocouple Input
Type K, 0 to 1600°F (-18 to 871°C)
(OS532, OS533, OS534 only)
Input Connection
SMP Connector
Thermocouple Display
±5°F (±3°C)
Accuracy @ 75°F (24°C)
Ambient Temperature
Thermocouple Display
Response Time
2 seconds
Operating Ambient
32°F to 122°F
Temperature:
(0°C to 50°C)
Operating Relative Humidity: 95% or less without condensation
Display:
Backlit LCD dual display
Keypad:
4 position, tactile feed-back membrane switch
7-1
Accuracy (24°C or 75°F
Ambient Temperature and at
emissivity of 0.95 or greater):
Summary of Contents for OMEGASCOPE OS523
Page 3: ...Unpacking Instructions n4 Notes...
Page 5: ...ii...
Page 15: ...General Description 1 1 8 Notes...
Page 63: ...Glossary of Key Strokes 8 8 2 Notes...
Page 73: ...Appendix Determining an Unknown Emissivity C C 4 Notes...