NetScan User’s Manual
Calibration 6-7
1.
Ensure the main unit is powered “OFF,” and has no power sources connected to it.
2.
If a signal conditioning card is in slot #1 of the chassis, remove the card.
3.
Install Calibration Card into Slot #1 (bottom slot) of the main unit.
&$87,21
Do not perform calibration until after the main unit and calibration equipment
have been powered on for at least one hour.
4.
With all covers in place, allow the main unit and calibration equipment to warm up (powered on) for at
least 1 hour.
5.
Place the Calibration Enable/Disable Pushbutton (on main unit rear-panel) in the “enable” (indent)
position to enable writing to RAM.
Note to
ChartScan users: Early production ChartScan units do not have a Calibration Enable/Disable Pushbutton.
For these units the calibration mode is enabled by positioning the DIP switch right-most micro-switch (#9) to the “1”
(up) position. Note that ChartScan units with the pushbutton have it located on the rear panel, as indicated in the
NetScan figure (page 6-2).
6.
If the calibration password has not been changed, enter
K12345
. If the password has been changed,
enter the K command and new password. This enables the calibration mode.
7.
Type
U2X
to verify calibration mode.
U128
should appear, indicating the unit is in calibration mode.
8.
Insert banana plugs of shorting option (shorting bar or 3 linked plugs) into -V, +V, and GND binding
posts of calibration card. This short circuits the analog backplane HI and LO voltage inputs and GND
(see previous figure, Calibration Card and Two Shorting Options).
9.
Type
H0X
. The Trigger LED will flash, indicating that offset calibration is in progress. When flashing
stops, the chassis offset calibration is complete.
10.
Type
U2X
.
U128
should appear, indicating the unit is in calibration mode and that no error has
occurred.
Response codes and meanings applicable to calibration are as follows:
Response to
U2X
Indication for Calibration Mode
U128
No error
U129
Read Failure (EEPROM Error)
U130
Write Failure (EEPROM Error)
U132
Chksum Err (EEPROM Error)
U136
Cal Err (offset, gain, or temperature
sensor out of range)
U144
Invalid password
U160
Invalid command
Summary of Contents for OMB-NETSCAN
Page 6: ...iv NetScan User s Manual...
Page 18: ...1 12 Configuring and Starting NetScan NetScan User s Manual Notes...
Page 38: ...3 8 General Information and Specifications NetScan User s Manual Notes...
Page 82: ...4 44 ChartView Software Reference NetScan User s Manual Notes...
Page 118: ...6 20 Calibration NetScan User s Manual...
Page 140: ...A ii NetScan User s Manual...
Page 192: ...API Command Reference Appendix A A 52 NetScan User s Manual Notes...
Page 237: ...Appendix D Registers Data Formats Queries NetScan User s Manual D 13...
Page 244: ...NetScan Program Examples Appendix E E 2 NetScan User s Manual...
Page 248: ...ASCII Code Summary Appendix F F 4 NetScan User s Manual Notes...
Page 250: ...NetScan Error Messages Appendix G G 2 NetScan User s Manual Notes...
Page 252: ...Abbreviations Appendix H H 2 NetScan User s Manual Notes...
Page 254: ...NetScan User s Manual...