Specifications
MOBILE WORKSTATION WM-DP1
29
The Olympus WM-DP1 mobile workstation and separation transformer is manufactured in the UK
by KeyMed (Medical & Industrial Equipment) Ltd.
Overload
protection
Input circuit breakers
2 x 13A
2 x 9A
Output fuse
T15AH
T10AH
Tighten to 0.5Nm reference
IEC 60127-6:1994
Potential
equalisation
Terminals fitted
2 at input, 2 x 6 at output
Classification
Standards compliance
The Separation Transformer conforms to the relevant
requirements of: EN IEC 60601-1/
EN IEC 60601-1-1:2001, Annex EEE/UL 60601-1 and
CAN/CSA Std. No. C22.2 No. 601.1-M90
Type of protection against
electric shock.
Class I equipment.
Degree of protection against
explosion
None
,
the Separation transformer should not be used
within the zone of risk of flammable gases
Degree of protection against
ingress of water.
IPX0 as defined in EN IEC 60529:1992, ordinary
equipment.
Mode of operation
Continuous
Environmental
conditions
Ambient temperature
Storage: -40
°
C to +70
°
C (-40
°
F to +158
°
F)
Operation: 10
−
40
°
C (50 - 104
°
F)
Atmospheric pressure
Storage: 23.5-106 kPa (3.4-15.37lbf/in
2
)
Operation: 70-106 kPa (10.15-15.37lbf/in
2
)
Relative humidity
95% maximum @ 40
°
C (104
°
F) non-condensing
Regulatory
status
European Economic Area
(EEA)
End of Life
220-240V units are CE marked in
accordance with Directive 93/42/EEC of
14 June 1993 concerning medical
devices, as an accessory to a medical
device, Class I. Year of manufacture is
given in first two digits of serial number.
In accordance with European Directive
2002/96/EC on Waste Electrical and
Electronic Equipment, this symbol
indicates that the product must not be
disposed of as unsorted municipal waste,
but should be collected separately. Refer
to your local Olympus distributor for return
and/or collection systems available in
your country.