UES-40
Inspection
4-2
ISSUE4
1. Preparation
Prepare the following equipment for inspection.
2. Inspection Procedure
Connect the Electrosurgical Analyzer to the UES-40. (Refer to the Electrosurgical Analyzer
instruction manual for details.)
Refer to the following table to check output values:
The above listed measurement points can cover all output specifications, because there are
no difference other than the amplitude of waveform in each output mode.
Name
Model
Specifications
Remarks
Electrosurgical Analyzer
454A
–
DNI NEVADA INC.
Table 4-1 Setting and output of each mode
Output mode
Load resistance [¾]
UES-40 output setting [W]
Standard [W]
Monopolar Spray
200
60
48-72
Monopolar Coag. 1
300
60
48-72
Monopolar Coag. 2
300
60
48-72
Monopolar Blend
300
125
100-150
Monopolar Cut
300
150
120-180
Bipolar Soft1
100
45
36-54
Saline Blend
100
160
128-192