UES-40
Product Specification
1-8
ISSUE4
3. Operating Requirements
(1) The operating conditions must be as follows.
Ambient temperature:
10°C to 40°C
Relative humidity:
30% to 85%
Atmospheric pressure: 700 hPa to 1060 hPa
Never use the UES-40 where flammable vapors occur.
(2) The power requirements are as follows.
(3) Do not use in combination with products not designated by Olympus.
(4) Before using the UES-40 for the treatment of patients wearing pacemakers, consult the
pacemaker manufacturer.
(5) Do not attach the patient plate for use near metallic implants.
(6) Do not wind UES-40 cords into loops or allow them to touch the patient or metallic parts
of the surgical table.
(7) Do not bundle UES-40 cords with cords of other medical equipment during use.
(8) Keep the UES-40 and its cords adequately separated from other medical equipment
and cords.
(9) If non-CF products are used in combination with the UES-40, do not use directly on the
heart.
(10) To ensure electrical safety, use of the following electronic medical equipment in combi-
nation with the UES-40 is prohibited:
a)
Equipment not yet proven safe when used with the UES-40
b)
Equipment not yet proven safe due to certain aspects of operation (such as the
leakage current)
(11) If the UES-40 is used in combination with endoscopic instruments for which the rated
voltage is listed in each instrument user manual, compare the open output-voltage char-
acteristics of the instruments with those of the UES-40 and adjust the output within the
USE-40’s rated voltage values. Refer to Fig. 1-1 to 1-4 for the open voltage characteris-
tics. If the user’s manuals for these instruments do not list the rated voltage, set the out-
put low initially and then gradually increase the output level.
Rated input voltage
120 V
220 V
Rated input current
12A
6A
Frequency
50/60 Hz
Permissible voltage fluctuation
±10%