6.5 Reprocessing the endoscopic CO2 regulation unit and accessories
61
UCR INSTRUCTION MANUAL
Ch.6
Cleaning
CAUTION
After wiping with a moistened lint-free cloth, dry the endoscopic CO
2
regulation unit
thoroughly before using it again. If it is used while still wet, there is the risk of an
electric shock.
4
Prepare another lint-free cloth moistened with
the surface disinfectant cleaner.
Figure 6.2
5
Disinfect the surfaces that previously contained patient debris (see Step 3) by wiping
with the moistened lint-free cloth.
6
Ensure that the surfaces are completely wet for the contact time instructed by the
surface disinfectant cleaner manufacturer.
7
If any surface of the endoscopic CO
2
regulation unit and accessories remains wet,
wipe it with a dry lint-free cloth and let it dry thoroughly.
1
Turn the endoscopic CO
2
regulation unit OFF and disconnect the power cord from the
hospital grade wall mains outlet.
2
Wipe with a dry lint-free cloth or a moistened lint-free cloth with water until dust and
dirt are removed.
3
If the surface of the instruments is wet, wipe it with a dry lint-free cloth and let it dry
thoroughly.
Lint-free cloth moistened with
the surface disinfectant cleaner
Summary of Contents for UCR
Page 2: ......
Page 5: ...Contents iii UCR INSTRUCTION MANUAL Specifications 69 EMC information 73...
Page 6: ...iv Contents UCR INSTRUCTION MANUAL...
Page 24: ...18 2 2 Front panel UCR INSTRUCTION MANUAL Ch 2 2 2 Front panel 3 1 2 4 5 6 7...
Page 26: ...20 2 3 Rear panel UCR INSTRUCTION MANUAL Ch 2 2 3 Rear panel 1 2 3 4 5 6...
Page 50: ...44 4 6 Inspection of the endoscopic CO2 regulation unit UCR INSTRUCTION MANUAL Ch 4...
Page 58: ...52 5 5 After use UCR INSTRUCTION MANUAL Ch 5...
Page 70: ...64 6 8 Transportation UCR INSTRUCTION MANUAL Ch 6...