3.3 Connecting a CO2 gas cylinder
27
UCR INSTRUCTION MANUAL
Ch.3
CAUTION
• If a significant gas leak is noticed from within the endoscopic CO
2
regulation unit,
stop using the endoscopic CO
2
regulation unit immediately and contact Olympus.
NOTE
Attach the CO
2
cylinder holder (MAJ-188, MAJ-1614) to the Olympus mobile
workstation (WM-NP1 or WM-WP1), and attach the CO
2
cylinder to the holder.
The cylinder hose must be attached to the endoscopic CO
2
regulation unit by Olympus personnel
during installation. The enclosed nut cover will be fixed during the installation when the cylinder hose
and the endoscopic CO
2
regulation unit are connected by Olympus personnel.
1
Inspect the cylinder hose for the endoscopic CO
2
regulation unit for damage, cracks,
and other irregularities.
2
Set the gas source selection switch on the rear panel to the “Cylinder” mode.
3
Use the supplied wrench to attach the cylinder
hose to the CO
2
gas inlet on the rear panel of
the endoscopic CO
2
regulation unit and
securely tighten it with a force of about
24.5 N·m (2.5 kgf/m) clockwise.
Figure 3.1
Cylinder hose
Wrench
Summary of Contents for UCR
Page 2: ......
Page 5: ...Contents iii UCR INSTRUCTION MANUAL Specifications 69 EMC information 73...
Page 6: ...iv Contents UCR INSTRUCTION MANUAL...
Page 24: ...18 2 2 Front panel UCR INSTRUCTION MANUAL Ch 2 2 2 Front panel 3 1 2 4 5 6 7...
Page 26: ...20 2 3 Rear panel UCR INSTRUCTION MANUAL Ch 2 2 3 Rear panel 1 2 3 4 5 6...
Page 50: ...44 4 6 Inspection of the endoscopic CO2 regulation unit UCR INSTRUCTION MANUAL Ch 4...
Page 58: ...52 5 5 After use UCR INSTRUCTION MANUAL Ch 5...
Page 70: ...64 6 8 Transportation UCR INSTRUCTION MANUAL Ch 6...