4.3 Inspection of the cylinder hose (MAJ-1080, MAJ-1081, MAJ-1082)
37
UCR INSTRUCTION MANUAL
Ch.4
WARNING
• If the cylinder hose is damaged, replace it with a new one.
• If the packing is damaged, replace it with a new one or replace the cylinder hose
with a new one.
• For information on how to replace the packing, refer to the leaflet “Replacement
procedure for packing (O-ring)”, which is included with the cylinder hose for UHI-3
(MAJ-1080).
WARNING
If the medical gas pipeline adapter is damaged, replace it with a new one.
Inspect the packing inside the clamp for scratches, cracks or other damage.
4.3
Inspection of the cylinder hose (MAJ-1080,
MAJ-1081, MAJ-1082)
1
Inspect the cylinder hose for scratches, cracks, or other damage.
2
Inspect the packing inside the clamp for
scratches, cracks or other damage.
Figure 4.1
4.4
Inspection of the medical gas pipeline adapter
(MAJ-1084, MAJ-1085)
Packing