background image

10

U-UVF248

}Before turning on the system, be sure to close the shutter by setting the shutter selector lever of the UV248 light source

box to the bottom position (marked {).

SUMMARY OF VISIBLE/UV LIGHT OBSERVATION PROCEDURE

Turn on the system.

Visible light observation

UV observation

(Operation)

1. Switch on the power supply for mercury/xenon burner (by setting the

power switch to “   ” and then setting the start switch to ON).

2. Switch on the microscope.
3. Switch on the TV camera/monitor for UV observation.
4. Switch on other set-up modules.

1. Set the light path of the reflected light illuminator to the BF position.
2. Engage a low-power objective for visible light observation in the light

path.

3. Place a specimen and adjust the focusing and brightness.
4. Engage a high-power objective for visible light observation and adjust

the focusing and brightness.

1. Set the light path of the reflected light illuminator to the DF position.
2. Engage the MApo100X objective for UV observation.
3. Set the shutter selector lever of the UV248 light source box to the upper

position (marked \).

4. Observing the monitor, adjust the focusing and brightness.

Summary of Contents for U-UVF248

Page 1: ...e Unit Model U UVF248 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this unit we recommend that you study this manual thoroughly before operating th...

Page 2: ......

Page 3: ...E 2 CONTROLS 3 PREPARATION 4 SUMMARY OF VISIBLE UV LIGHT OBSERVATION PROCEDURE 5 SPECIFICATIONS 6 TROUBLESHOOTING GUIDE FOR UV OBSERVATION 7 ASSEMBLY 5 6 7 8 9 10 11 12 13 16 U UVF248 1 Centering the...

Page 4: ...irradiate human eyes or skin 2 To prevent unexpected irradiation of UV rays be sure to close the idle objective positions of the revolving nosepiece by attaching the caps provided with the revolving...

Page 5: ...e to cool down the burner 10 The top panel of the lamp housing becomes very hot during operation To prevent a fire hazard be sure not to block ventilation of this part A space of 10 cm or more should...

Page 6: ...del BX URA2 Provided that an idle mirror unit or the U MDF 3 DF mirror unit is used 3 Objectives UV observation MApo100X objective for 248 nm UV rays only NA 0 9 WD 0 2 mm Actual field of view 50 m Vi...

Page 7: ...lations and rules of your local government and be sure to observe them The used mercury xenon burner should be disposed of as an industrial waste If you cannot dispose of it by yourself contact Olympu...

Page 8: ...V camera Sony XCD SX910UV Image measuring software analySIS FIVE Active AF unit Reflected Light Illuminator BX RLA2 BX RLAA BX URA2 UV Objective MApo100X for 248 nm UV Rays Objectives for visible ligh...

Page 9: ...13 TV adapter clamping screw Page 13 Caution label Page 1 The illumination lens adjustment screw should be left in the factory shipped setting position closer to the UV light guide mount It needs to...

Page 10: ...using knob Page 8 Burner centering knobs Page 8 Burner socket Page 14 UV Light Guide U UVF2FB U UVF5FB Output connector Page 15 Input connector Page 15 Power Supply for Mercury Xenon Burner Mfd by USH...

Page 11: ...the centering scope by turning the cap counter clockwise 4 Observing the UV cut centering scope adjust the following controls so that the brightest and most well focused part of the arc image comes on...

Page 12: ...n into focus by observing the monitor for UV observa tion If the observation image is not displayed or abnormally dark the illumination center is deviated In this case the light guide mount should be...

Page 13: ...camera monitor for UV observation 4 Switch on other set up modules 1 Set the light path of the reflected light illuminator to the BF position 2 Engage a low power objective for visible light observat...

Page 14: ...D size of 2 3 inch or less Recommended model Sony XCD SX910UV 1 2 inch CCD Dimensions weight 108 W x 157 5 H x 259 D mm approx 1 9 kg 248 Light Source Box U UVF248LB Item Specifications Applicable lig...

Page 15: ...f the light source box is closed the lever is set to the position marked Set the shutter selector lever to the upper position marked to open the shutter The light path of the reflected light illu mina...

Page 16: ...be into the observation tube mount and tighten the observation tube clamping screw When Using the Active AF Unit Mount the AF unit above the reflected light illuminator With certain AF unit models the...

Page 17: ...e burner is replaced Remove the holder or old burner by loosening the two burner holding screws 4 Attach the burner so that the negative pole comes at the top and tighten the negative pole side with o...

Page 18: ...amping knob is aligned with the positioning reference plane Attach the objective in the same way as a UIS2 UIS objective To prevent unexpected irradiation of UV rays be sure to close the idle objectiv...

Page 19: ...nstruction manual provided with the power supply for mercury xenon burner The following description pertains to the settings required for use of this unit and the method of attaching the conversion ca...

Page 20: ...inspections are charged after expiration of the warranty period If you have any questions please contact Olympus Check results Date Check items 1 More than 8 years have passed since original purchase...

Page 21: ...MEMO...

Page 22: ...MEMO...

Page 23: ......

Page 24: ...AX7645 03 Manufactured by Distributed by...

Reviews: