Applications 117
6.
Press the SELECT
menu soft key
, this will log the next point/echo. SELECT will
automatically change to ADJUST. With the UP/DOWN arrow keys or SmartKnob™,
adjust the gain to increase the echo amplitude to the desired height: for example, from
60% FSH to 80% FSH.
7.
Continue following the above steps for all the points/echo that require DAC (see
Figures
5-61 and 5-62
for the points/echo at the same amplitude).
Figures 5-60 - 62