EN-4
OmniScan X3 — Getting Started Guide
The OmniScan X3 DC power adaptor (P/N: OMNI-A-AC [U8767093]) is designed to power the
OmniScan X3 and charge the lithium-ion batteries only (P/N: OMNI-A-BATT [U8760010]). Do not
attempt to charge any other type of battery, or use any other chargers/adaptors (except those
recommended by Olympus—for example, the optional external battery charger, P/N: OMNI-A-CHRG
[U8767077]). Doing so might cause an explosion or injury. Do not attempt to power or charge other
electronic equipment with the DC power adaptor (P/N: OMNI-A-AC [U8767093]) or external battery
charger (P/N: OMNI-A-CHRG [U8767077]), unless specifically instructed in the manual. Misuse of the
DC power adaptor or external battery charger can cause other batteries and/or instruments to explode,
which could lead to serious injury or death.
Connect the OmniScan X3 flaw detector to the DC power adaptor that is connected to an
appropriate AC power source.
OR
Insert a battery into the optional external battery charger (P/N: OMNI-A-CHRG [U8767077]) that is
connected to an appropriate AC power source.
Battery charging in the OmniScan X3 flaw detector is performed as follows:
— When the OmniScan X3 is turned off:
When the OmniScan X3 is connected to a proper DC power source and is turned off, it will
automatically recharge the battery.
The power indicator light flashes orange to indicate that the battery is currently being recharged.
When the charge is complete, the power indicator light emits a steady orange light. The battery
can take up to 3 hours to fully charge from a completely discharged state (less than 5 %
remaining charge).
— When the OmniScan X3 is running:
When the OmniScan X3 is running and connected to a proper DC power source, it will
automatically recharge the battery. The power indicator light is green during charging (when
running).
Because the OmniScan X3 is running, less power is available from the DC source to recharge the
battery. As such, it could take up to 8 hours to charge a completely discharged battery.
Battery charge status indicators in the software interface are shown below.
WARNING
Missing or incorrectly installed battery
Fully charged (disconnected from DC power adaptor)
Battery level (combined remaining charge percentage)
Level increments in 10 % steps (0–100 %)
Charging (interior blinking) with percentage of charge attained
Fully charged (connected to DC power adaptor)
Too hot to charge
Too hot to operate, or critical temperature (rapid blinking)