OmniScan MX2 — Getting Started Guide

Figure 1  Battery replacement

The OmniScan MX2 can operate using only one battery, making it possible to remove the other battery 
while the unit is in operation (hot swap). When two batteries are used, the OmniScan MX2 
automatically discharges both batteries at the same time, increasing the remaining operating time by 
about 10 %. To discharge one battery at a time, simply remove the other battery.

Charging the Batteries

To charge the batteries

Connect the OmniScan MX2 to the supplied DC power adaptor.

When the OmniScan MX2 is connected to a proper DC power source and is turned off, it 
will automatically recharge any battery requiring charging, starting with the weakest 
battery (if the recharge conditions are met).
Once the weakest battery has reached the level of the other battery, both batteries are 
charged simultaneously.
The power indicator light indicates that batteries are currently being recharged when it 
flashes orange. When the charge is complete, the power indicator light emits a steady 
orange light. Each battery can take up to 3.5 hours to fully charge from a completely 
discharged state (less than 5 % remaining charge).

When the OmniScan MX2 is running and connected to a proper DC power source, it will 
automatically recharge any battery requiring charging, starting with the weakest battery (if 
the recharge conditions are met). The indicator for the battery being recharged flashes 
yellow.
Once the weakest battery has reached the same level as the other battery, both batteries are 
charged simultaneously.
Because the OmniScan MX2 is running, less power is available from the DC source to 
recharge the battery or batteries. As such, it could take up to 8 hours to charge each 
completely discharged battery. For more information about battery charge status, refer to 
the 

OmniScan MX and MX2 User’s Manual

.

The OmniScan MX2 DC power adaptor (P/N: OMNI-A-AC [U8767093]) is designed to charge 
OmniScan MX2 batteries only (P/N: OMNI-A-BATT [U8760010]). Do not attempt to charge any other 
type of battery, or use any other chargers/adaptors. Doing so could cause an explosion or injury. Do 
not attempt to power or charge other electronic equipment with the OMNI-A-AC DC power adaptor, 
as doing so could cause an explosion resulting in death or serious personal injury.

Battery Status Indicators

When the OmniScan MX2 is running, the battery status is shown by indicators located on the upper-
left corner of the touch-screen display. These indicators display the amount of power remaining in 
each battery using one of the two following methods:

The remaining operational time is displayed inside the battery status indicator. This 
information can only be displayed accurately after the OmniScan MX2 has been operating 
for approximately 15  minutes.

Summary of Contents for OmniScan MX2

Page 1: ...also available on the documentation USB drive that is shipped with the OmniScan MX2 or can be downloaded at www olympus ims com Safety Signal Words Indicates an imminently hazardous situation calling...

Page 2: ...15 V can be present on the PA connector The warning symbol near the PA and UT connectors signals this electric shock risk Alarm and I O connector Scanner interface SVGA video output Ethernet RJ 45 por...

Page 3: ...n MX2 or an external charger approved by Olympus to charge the battery Do not recharge a battery unless the indicators light up when the capacity check button on the battery is pressed Doing so could...

Page 4: ...y requiring charging starting with the weakest battery if the recharge conditions are met The indicator for the battery being recharged flashes yellow Once the weakest battery has reached the same lev...

Page 5: ...n both batteries are used simultaneously they will both appear highlighted The OmniScan MX2 software keeps the user informed of the battery status The table below describes some of the battery status...

Page 6: ...control area alone Element Name Description Scroll knob Used to navigate through selections without the need for a keyboard mouse or touch screen Turning the scroll knob in a clockwise direction shift...

Page 7: ...message appears 2 Select Shut Down A message asking you if you want to save your setup appears 3 To save your setup select Yes You can also turn off the OmniScan MX2 by pressing and holding the Power...

Page 8: ...olland Hitech50 which contains 50 post consumer fiber DMTA 20019 01EN Rev B February 2013 DMTA 20019 01EN 50 OmniScan MX2 Getting Started Guide Adaptors The figure below shows how the OmniScan MX2 can...

Reviews: