20
DIP Switch Setting (Fig. 30)
# Confirm the correct connections before proceeding.
}Set the main switch of the microscope to “ ” (OFF) before setting the
DIP switch @.
The DIP switch settings are read and entered at the moment the
microscope is turned ON.
# Be careful not touch the internal circuit boards when setting the DIP
switches. Otherwise, the internal circuitry may be damaged by static
electricity. Since the human body is charged with a small amount of
static electricity, discharge it from your body before proceeding to
the setup. This is done by lightly touching any metallic object.
Function
Setting Result
Fig. 30
14
(Fig. 30)
Setting the DIP Switches and
PRESET Switches
DIP Switches
1
2
3
4
Revolving
nosepiece type
(Number of holes)
Aperture iris diaphragm
control button setting
Glare prevention during
objective switching
(Darkfield observation)
U-D5BDREMC/P5REMC
U-D6REMC
Setting enabled.
Setting disabled.
Glare prevention
No glare prevention
OFF
ON
OFF
ON
OFF
ON
: Factory setup.