MV-1/MV-2 Endoscope Disinfectors
Frequently
Asked
Questions
09FQ0015 6/18
Two Corporate Center Drive, Melville, NY 11747
Copyright 2001, Olympus America Inc.
Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part
thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do
Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.
➨
Caution:
If the electrical outlet is adjacent to the water supply, a ground-fault circuit interrupter
(GFCI) should be used.
➨
Plug one end of the power cord securely into the power entry module on the right side of the MV-
1/MV-2.
➨
Plug the other end of the power cord into a standard 110 VAC outlet.
4. MV-1/MV-2 Drain Screen/Drain Cover Installation
#
Drain Screen/Drain Cover
➨
Remove the packaging label from the drain screen and make sure the drain screen is in place.
➨
Place the white cover over the drain screen. It should fit behind the drain bar.
➨
Caution:
Do not attempt to lift up the MV-1/MV-2 using the drain bar. Damage to the MV-1/MV-2
may result.
5. Connection of the Channel Connectors to the MV-1/MV-2. Refer to the
for a listing of endoscopes and their corresponding channel connectors.
#
Channel Connectors
➨
Attach the channel connectors to the quick-release connectors in the MV-1/MV-2 basin. Be sure
to press them in until they click into place.
6. Accessory Bag Placement
#
Accessory Bag Placement
➨
Place the accessory bag into the MV-1/MV-2.
➨
Caution:
Do not place the MV-1/MV-2 accessory bag over the drain screen as it might impede
thorough draining.
➨
Caution:
Do not hang the accessory bag over the liquid-level switch as it might impair proper
level measurement.
7. Connection of the Water Filtration System (CLM-1076) to the Water Supply
#
Water Filtration System (CLM-1076)
➨
Note:
In order to comply with the recommendations of your hospital’s infection control committee,
and national and local guidelines and policies, it may be necessary to disinfect or sterilize the MV-
1/MV-2 and the water filtration system. Instructions for this are provided in
8.19 of document # 09-WI-0239.
Also see the
MV-1/MV-2 Endoscope Disinfectors User/Service
Manual.