Olympus MV-1 Frequently Asked Questions Manual Download Page 2

MV-1/MV-2 Endoscope Disinfectors

Frequently

Asked

Questions

09FQ0015 2/18

Two Corporate Center Drive, Melville, NY 11747

Copyright 2001, Olympus America Inc.

Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part

thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do

Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.

Model Overview

Features

  An optional mobile cart (Item CLM-1005) provides a station for disinfector mounting, storing of related

accessories, disinfectant reservoir, and optional water filtration system.

  The compact size of the disinfectors saves space.

  All fluids are pumped continuously during fluid cycles.

  All fluid cycles are followed by an air purge cycle to clear internal tubing and channels of all fluids.

  Complete disinfection of all internal fluid tubing is provided.

  Each channel connector has an air pump.

  Two fresh water rinses follow the high-level disinfection cycle to provide maximum rinsing.

  The high-level disinfectant reservoir allows visual inspection of the disinfectant, easy access for disinfectant

testing, and is removable for cleaning.

  The disinfectors are built without open vents and operate as closed systems to allow minimal fume

emission. Data reports that less than 0.04ppm glutaraldehyde vapor is emitted during normal operation.

  Disinfectant can be pumped directly into and out of the reservoir using an automated transfer pump. This

reduces operator exposure, and allows disinfectant disposal directly into containers, instead of into a drain.

Note:

Item CLM-1087 is a recommended LCG transfer pump.

  The disinfectors allow for installation of a water filtration system, which is recommended.

Note:

Item CLM-1076 is a recommended two-stage water filtration system which provides 1 micron at pre-
filter and 0.2 micron at final filter.

Models

 

MV-1

Disinfects one flexible endoscope per cycle and is equipped with four individual endoscope
channel connectors. Each channel connector is connected directly to the pump without “Y” or
“T” fittings. This results in maximum flow through channels regardless of the channel size.

 

MV-2

Disinfects two flexible endoscopes in a single cycle. The unit is equipped with seven channel
connectors, enabling the connection of each endoscope channel to a dedicated connector.

Summary of Contents for MV-1

Page 1: ...tronic or otherwise without the expressed written consent of Olympus America Inc is strictly prohibited Only printed copies with Controlled Document Do Not Copy stamped in red are considered controlled copies All other printed copies are considered Uncontrolled and may not contain the latest document revisions Troubleshooting Rinse Alarm Model Overview Cabling and Connectivity Compatible Component...

Page 2: ...r rinses follow the high level disinfection cycle to provide maximum rinsing The high level disinfectant reservoir allows visual inspection of the disinfectant easy access for disinfectant testing and is removable for cleaning The disinfectors are built without open vents and operate as closed systems to allow minimal fume emission Data reports that less than 0 04ppm glutaraldehyde vapor is emitte...

Page 3: ... other printed copies are considered Uncontrolled and may not contain the latest document revisions Model Overview Cont d Additional Components The following components are supplied with the MV 1 MV 2 Endoscope Disinfectors MV 1 MV 2 User service manual Accessory bag 1 Alcohol flushing adapter Water hose Drain hose Power plug Endoscope connector standard set for Olympus or Pentax or Fujinon flexib...

Page 4: ...of Olympus America Inc is strictly prohibited Only printed copies with Controlled Document Do Not Copy stamped in red are considered controlled copies All other printed copies are considered Uncontrolled and may not contain the latest document revisions Compatible Components The MV 1 MV 2 Endoscope Disinfectors are compatible with the following components Mobile Cart CLM 1005 LCG transfer pump CLM...

Page 5: ...1 Connection of the Drain to the MV 1 MV 2 Flexible drain hose Connect the flexible drain hose to the white drain fitting on the back of the MV 1 MV 2 and attach the screw clamp Caution The drain must be 12 in below the drain fitting on the disinfector and loops in the drain hose must be eliminated to ensure that fluids drain thoroughly Incomplete draining will result in dilution of the LCG and or...

Page 6: ...lace Place the white cover over the drain screen It should fit behind the drain bar Caution Do not attempt to lift up the MV 1 MV 2 using the drain bar Damage to the MV 1 MV 2 may result 5 Connection of the Channel Connectors to the MV 1 MV 2 Refer to the MV 1 MV 2 Connector Matrix for a listing of endoscopes and their corresponding channel connectors Channel Connectors Attach the channel connecto...

Page 7: ...e hose fitting water inlet on the back of the MV 1 MV 2 9 Installing the Filters in the Water Filtration System CLM 1076 Filters Prefilter CF01 0017 Postfilter CF01 0018 or Prefilter and Postfilter Set CF01 0007 Caution Personal protective equipment should be worn to guard against reprocessing chemicals and potentially infectious material when performing this procedure Wear appropriate equipment s...

Page 8: ...of the filter and are centered when the filter is inserted into the housing If the postfilter has been removed place the new postfilter into the blue postfilter housing with the black O ring facing UP and the closed end facing DOWN Make sure the O ring is securely fastened to the peg at the top of the housing It may be necessary to moisten the O ring with water or a water based lubricant 10 Connec...

Page 9: ...rs User Service Manual All endoscopes must be precleaned before placement in the MV 1 MV 2 Follow the manufacturer s instructions and established professional guidelines see page 5 in the MV 1 MV 2 Endoscope Disinfectors User Service Manual for further instructions on the cleaning steps that must be performed prior to placing an endoscope into the MV 1 MV 2 Caution Effective reprocessing can only ...

Page 10: ...parts and filters must be used to assure proper operation and efficiency of the MV 1 MV 2 Do not use gauze cotton swabs or other lint producing cleaning aids on the MV 1 MV 2 These materials shed fibers that can become caught in the channels of endoscopes or in the components of the MV 1 MV 2 Do not use products that contain alcohol when cleaning the lid of the MV 1 MV 2 Alcohol may cause crazing ...

Page 11: ... 1 MV 2 Water Flow A channel connector should be connected to each port in the MV 1 MV 2 basin Turn the MV 1 MV 2 power switch ON The power switch is located on the right side of the MV 1 MV 2 Press START on the MV 1 MV 2 control panel Wait for the pump to feed water through the channel connectors Confirm that water is flowing properly through each connector Note Flow volume will vary depending up...

Page 12: ...nimum of 4 gallons 15 1 L of LCG should be in the LCG reservoir of the MV 1 A minimum of 5 gallons 18 9 L of LCG should be in the LCG reservoir of the MV 2 Note Recommendations for the disposal of LCG vary between formulations and locations Follow the policies and procedures applicable to the facility Before Operating LCG Concentration The LCG must be tested daily to ensure that it contains an ade...

Page 13: ...he system to operate properly The MV 1 MV 2 disinfectors require a minimum incoming water pressure of 40 psi 2 8 bar as indicated by the reading on gauge 1 Check the gauges located on the water filtration system When the reading on gauge 1 is normal and the reading on gauge 3 drops below 40 psi it is likely that one or both of the filters must be changed Read the gauges to determine which filter m...

Page 14: ...n occur and possible solutions to these problems Some of the solutions given may require the intervention of a trained service engineer If this is the case contact Olympus The following documents are referenced in this section Click on the document title below to open the document MV 1 MV 2 Endoscope Disinfectors User Service Manual Document 09 WI 0240 MV 1 MV 2 Endoscope Disinfectors Confidence T...

Page 15: ...bleshooting LCG 1 Was the LCG lost through the overflow drain Check to see if the MV 1 MV 2 is level If necessary adjust the legs of the MV 1 MV 2 Make sure any material or debris is not covering the float tower Make sure that the float sensor is not defective Replace the float sensor if necessary 2 Is the LCG not completely filling the basin Make sure the LCG tubing does not contain any kinks or ...

Page 16: ...h If drainage is too slow contact Olympus Check the Solution Out solenoid to be sure it is not leaking water This applies only to the LCG reservoir overfilling problem If there is water leakage contact Olympus 5 Is there loss of LCG during MV 1 MV 2 operation If LCG is lost via the drain contact Olympus Check the drain screen to ensure that the drain is not blocked Make sure the LCG OUT quick conn...

Page 17: ...fully inserted Ensure the check valve that is in line with the alcohol manifold is not leaking air If this is the case replace the check valve Make sure the air pump is not defective If necessary replace the air pump Check the internal air pump tubing to be sure it is not loose If necessary reconnect the tubing properly If air is flowing instead of fluid the motherboard might be defective If neces...

Page 18: ...ed and may not contain the latest document revisions Troubleshooting MV 1 MV 2 Control Panel Cont d Ensure the MV 1 MV 2 control panel is operating properly If necessary replace the MV 1 MV 2 control panel Check the ribbon cables within the MV 1 MV 2 to be sure that they are connected properly and are not loose Troubleshooting Returning the Unit for Repair If the MV 1 MV 2 Endoscope Disinfectors m...