 2  General safety instructions 

9

2.3

Avoiding device damage     

Never open the camera housing!

The  camera's  CMOS  chip  is  extremely  sensitive.  Electrostatic  discharge  can
permanently damage the camera. 
Therefore, never, under any circumstances, open the camera housing!

Avoid vibration shocks!

Mechanical shocks and strong vibrations can damage the camera. 
Don't drop the camera during installation. Make sure it is securely mounted on the
microscope. When it's not in use, store it in a safe place!

Secure microscopes with additional components!

A top-heavy microscope can tip over or fall down and result in damage to both the
camera and the microscope. 
If the microscope is top-heavy or otherwise unstable, fix it to the wall!

Don't touch the protective glass!

The protective glass over the CMOS chip is extremely sensitive and must not come
into contact with bare hands or any other objects, otherwise the acquisitions will be
impaired by fingerprints or scratches. 
Remove  dust  from  the  protective  glass  carefully  with  compressed  air.  For  the
removal of dust stuck due to static electricity, ionized air is recommended.

Observe the operating conditions!

Exposure to high temperatures or humidity can damage component parts of the
camera. 
Always observe the operating and storage conditions (see chapter 1.4 on page 6).

Don't expose the camera to strong light!

Exposure  to  strong  light  can  cause  the  camera  to  overheat,  which  may  cause
damage to its component parts. 
Do not expose the camera to strong light (such as sunlight) for long periods of time.

Only use the USB cable that is delivered with your camera

Operate the camera only with the USB cable that is part of the camera's scope of
supply. This ensures that the EMC basic standards are fulfilled.

ATTENTION

ATTENTION

ATTENTION

ATTENTION

ATTENTION

ATTENTION

ATTENTION

Summary of Contents for LC30

Page 1: ...English Optical Microscope Accessory To obtain optimum performance of this camera and to ensure the safety study this manual thoroughly before operating the camera and keep it on hand during operation...

Page 2: ...ndirect special incidental or consequential damages arising out of purchase or use of this manual or the information contained herein No part of this document may be reproduced or transmitted in any f...

Page 3: ...oiding personal injuries 8 2 3 Avoiding device damage 9 3 System environment 10 3 1 General remarks 10 3 2 Microscopes and camera adapters 12 3 3 PC 13 3 4 Software 14 4 Installation 15 4 1 Step 1 Ins...

Page 4: ...e cap Camera with protective cap USB 2 0 cable USB 2 0 cable Installation Manual Installation Manual USB 3 0 PCI express card with driver on CD and power supply cable not included in the scope of supp...

Page 5: ...ope by a C mount The ICC profile technology guarantees an optimal color mapping Binning modes The OLYMPUS LC30 has three binning modes In binning mode the camera s CMOS sensor combines neighboring pix...

Page 6: ...tions The maximum permissible range for the ambient temperature during camera operation is 5 C to 40 C The maximum permissible relative humidity during camera operation is 80 condensation free When th...

Page 7: ...re and during the installation 2 1 Explanation of the symbols used Personal injuries Device damage General notes The warning sign and the word WARNING indicate dangerous situations that can lead to se...

Page 8: ...In the case of a fire the material of the camera can give off toxic gases When these are inhaled it can cause irritation of the respiratory passages and damage to them Wear respiratory protection when...

Page 9: ...ensitive and must not come into contact with bare hands or any other objects otherwise the acquisitions will be impaired by fingerprints or scratches Remove dust from the protective glass carefully wi...

Page 10: ...30 can be used in combination with many available micro scopes However its performance in combination with non Olympus microscopes has not been tested Application context Explanation Countermeasure Us...

Page 11: ...y closed Switch on the shading correc tion or decrease illumination intensity and open the AS diaphragm Observing high contrast samples High contrast samples may cause the auto exposure to fail result...

Page 12: ...era adapters with the magnification factor 0 35 or larger are suitable U CMAD3 U TV1x 2 U TV0 5xC 3 U TV0 63xC GX TV0 5xC GX TV0 7xC MVX TV0 63xC MVX TV1xC Upright microscopes BX61 BX51 BX41 CX41 CX31...

Page 13: ...the camera with a laptop Generally speaking it is possible to operate the camera with a laptop If the power supply is not sufficient to operate the camera when an integrated USB interface is used conn...

Page 14: ...dardized software interface between software programs and image acquisition hardware such as digital cameras or scanners You can download the DP2 TWAIN software from this Olympus website https www oly...

Page 15: ...acturer Also pay attention to the safety instructions given below After you have inserted the card make sure to note the following point This step is only necessary when a USB PCI express card is incl...

Page 16: ...programs that run in the background for example antivirus software 2 Browse to the directory that contains the software and double click the setup exe file The installation wizard starts Follow the i...

Page 17: ...cable with the type B plug into the camera socket Make sure that you only use the USB cable that is delivered with your camera The proper functioning of the camera can only be guaranteed when this cab...

Page 18: ...allation only takes a few seconds Please pay attention to the icons displayed in the status bar 5 Wait until the installation of the camera driver has been finished The camera has now been registered...

Page 19: ...an gauze moistened with pure alcohol Cleaning the housing Parts other than the glass components should be cleaned by wiping with a clean cloth Do not use organic solvents to remove major stains Use a...

Page 20: ...tions Install or maintain imaging software Maintain imaging software Provide licenses update installation When the Image source selection dialog box is displayed select the Olympus manufacturer and se...

Page 21: ...h the European Union directive on the restriction of the use of certain hazardous substances in electrical and electronic equipment 2011 65 EU 7 4 FCC conformity USA This device complies with part 15...

Page 22: ...eorient or relocate the receiving antenna 2 Increase the separation between the equipment and receiver 3 Connect the equipment into an outlet on a circuit different from that to which the receiver is...

Page 23: ...7 Notes on conformity and disposal 23 7 6 China RoHS conformity China...

Page 24: ...63 0914 Japan OLYMPUSEUROPASE CO KG Wendenstrasse 14 18 20097 Hamburg Germany OLYMPUSCORPORATIONOFTHEAMERICAS 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A OLYMPUSCORPORATIONOFA...

Reviews: